QC - THE REGULATIONS
Sharon S. Ehrmeyer, Ph.D.
If you perform laboratory tests in the US, the government's
got you covered! One way or another, you have to comply with the
Clinical Laboratory Improvement Amendments of 1988 (CLIA88),
which establish the minimum standards for all laboratory testing,
including specific regulations for quality control [1]. You can,
however, select the organization that will administer these regulations
or administer professional standards that are equivalent to these
regulations:
- The Health Care Financing Administration (HCFA) will inspect
any size laboratory, including physician office laboratories,
for adherence to the CLIA standards;
- The Commission of Office Laboratory Accreditation (COLA)
inspects only physicians office laboratories for adherence to
COLA standards which closely parallel the CLIA regulations;
- The Joint Commission on Accreditation of Healthcare Organizations
(JCAHO) inspects laboratories as part of an overall inspection
of a hospital or healthcare organization;
- The College of American Pathologists Laboratory Accreditation
Program (CAP-LAP) mainly inspects large laboratories directed
by pathologists.
There are other government approved organizations that have
standards for laboratories to follow and some states impose specific
requirements [2].
Health Care Financing Administration
(HCFA) and Commission of Office Laboratory Accreditation (COLA)
Requirements for QC
The first part of this discussion focuses on CLIA88 and
COLA regulations [1,3] since COLA's 200 question checklist adheres
to the CLIA88 QC requirements almost exactly. Professional
standards administered via CAP and JCAHO are considered later.
CLIA QC Requirements based on Test Complexity
CLIA'88 regulations are based on four categories of test complexity:
waived, provider performed microscopy (PPM), moderate complexity,
and high complexity. Current information on test complexity can
be obtained from the CDCs web site (http://www.cdc.gov/phppo/dls/testcat.htm).
Each testing category has different regulatory requirements for
personnel, quality control, quality assurance, proficiency testing,
etc.
Waived testing
This category requires the least regulation. Under CLIA and
COLA, the minimum requirement for anyone performing waived testing
is to follow the manufacturers directions for QC, and if
no directions are included, to follow good laboratory practices.
No specific QC requirements are identified. Good laboratory practice
would dictate that controls be run and results documented and
reviewed for correctness before reporting patient results. The
tests that are waived include the following (current September
29, 1997 list of Waived Tests):
- Dipstick or tablet reagent urinalysis for bilirubin, glucose,
hemoglobin, ketones, leukocytes, nitrite, pH, protein, specific
gravity, and urobilinogen
- Fecal occult blood
- Ovulation tests - visual color comparison tests
- Urine pregnancy tests - visual color comparison tests
- Erythrocyte sedimentation rates - non-automated
- Blood glucose - all glucose monitoring devices cleared by
FDA for home use; HemoCue B-Glucose System, and Cholestech L.D.X.
- Hemoglobin - (non-automated) by copper sulfate; (automated)
by HemoCue Hemoglobin System
- Hematacrit - all spun microhematocrit procedures, Wampole
STAT-CRIT, Separation Technology STI HemataSTAT II and Model
C70, StatSpin Technologies CritSpin, and Vulcon Technologies
Microspin 24, Protime - ITC (International Techidyne Corp.) and
BMC CoaguChek
- H. pylori -- Serim Pyloritek Test Kit, GI Suppy HP-FAST,
Quidel QuickVue One-Step for whole blood, Delta West CLOtest,
Abbott FlexPack HP, ChemTrak AccuMeter, and SmithKline FlexSure
- Strep A (throat only) Quidel QuickVue In-Line One-Step, Binax
NOW, Abbott Signify and SmithKline ICON Sx
- Cholesterol -- Cholestech L.D.X., BMC Accu-Chek InstantPlus,
ChemTrak AccuMeter, and J & J ADVANCED CARE
- HDL Cholesterol -- Cholestech L.D.X.
- Triglyceride -- Cholestech L.D.X
- Gastric occult blood -- SmithKline Gastroccult
- Micro albumin -- BMC Chemstrip Micral
- All qualitative color comparison pH testing for body fluid
(other than blood)
To keep up to date with all the new test additions and new
methodologies for waived tests, you can check the CDC
Web site [http://www.cdc.gov/phppo/dls/clia.htm].
Provider Performed Microscopy (PPM)
As of January 1993, the category "Physician-Performed
Microscopy" was established under CLIA88. In the April
24, 1995 Federal Register, the category was renamed to
"Provider-Performed Microscopy." This category is a
subset of moderate complexity and is exclusively for physicians,
dentists, nurse practitioners and midwives, and physician assistants
performing the testing as part of a patient examination. The primary
instrument for performing the test(s) is the microscope.
The 9 tests identified under a PPM CLIA certificate can be
performed, as well as waived tests. When these tests are provided,
the practitioners are expected to follow the manufacturers
directions for QC or follow good laboratory practices. Good laboratory
practice would dictate that controls be run whenever possible
and results documented and reviewed for acceptability before reporting
patient results. The PPM category includes:
- all direct wet mount preparations (suspended in saline or
water) for the presence or absence of bacteria, fungi, parasites,
and human cellular elements
- KOH preps
- pinworm exams
- fern tests
- post-coital direct qualitative exams of vaginal or cervical
mucous
- urine sediment exams
- nasal smears for granulocytes
- fecal leukocyte exams
- qualitative semen analysis (limited to the presence or absence
of sperm and detection of motility).
Moderate Complexity Tests
Approximately 75% of all tests performed in laboratories today
fall in the category of moderate complexity. For this category,
the QC requirements are identified in subpart K for CLIA88
-- Quality Control. Originally, in the February 28, 1992 Federal
Register, HCFA stated that as of September 1, 1994, laboratories
would meet most of CLIA'88 QC requirements simply by following
manufacturers' labeling, provided the FDA approved manufacturers'
QC instructions. This never happened and the implementation date
(9/1/94) has now been suspended, in the May 12, 1997 Federal
Register, until at least 7/31/98.[5] Until HCFA implements
new QC requirements, laboratories performing moderate complexity
tests will meet the CLIA QC requirement by following sections
§493.1201 and §493.1202(c) only.
Section §493.1201 (general QC: moderate or high complexity
testing, or both) states: The laboratory must establish and
follow written QC procedures for monitoring and evaluating the
quality of the analytical testing process of each method to assure
the accuracy and reliability of patient test results and reports.
Section §493.1202(c) identifies the specific requirements
for unmodified moderate complexity tests: [1,4]
- follow manufacturer directions;
- have a procedure manual for the method that identifies how
to perform the testing and reporting of results;
- perform and document calibration procedures or check calibration
at least once every six months;
- assay at least two levels of control materials each day of
testing (a run cannot exceed 24 hours) and keep records;
- perform and document any applicable specialty and subspecialty
control procedures;
- perform and document remedial actions as specified in §493.1219;
- maintain records of all QC activities for 2 years or 5 years
for immunohematology and 10 years for pathology as specified
in §493.1221.[4]
As you can see, CLIA QC includes a lot more than traditional
statistical QC and defines standards for calibration, procedure
manuals, remedial actions and record keeping. Concerning statistical
QC, for most moderate complexity tests, the general requirement
is to analyze two levels of QC materials on each day of testing.
However, for certain tests, i.e., blood gases, hematology and
coagulation tests, etc., CLIA requires additional QC:
- Blood gases require, at a minimum, one control sample every
8 hours of testing and a calibrator or control in each run unless
the instrument "autocals" at least every 30 minutes.
- Automated hematology and coagulation test systems require
two levels of controls every 8 hours of testing and each time
a change in reagent occurs.
- Manual cell counts using a hemocytometer must be tested in
duplicate and one control is required every 8 hours of operation.
- For manual coagulation testing, each analyst must perform
two levels of controls before testing patient samples and with
each change in reagent. In addition, patient and control samples
must be tested in duplicate.
- With electrophoresis, one control needs to be included in
each electrophoretic cell and the control must contain fractions
representative of those routinely reported in patient samples.
- For toxicology, each thin layer chromatogrophy (TLC) plate
must be spotted with at least one sample of calibration material
containing all drug groups reported by the laboratory with TLC,
one control must be included in each chamber, and the sample
must be processed through each step of patient testing including
extraction.
- Qualitative tests with built-in controls are adequate provided
the kit has been qualified with at least one positive and negative
control. CAP requires one positive (external) and one negative
(external) control per day.
- For those methodologies where the manufacturer specifies,
surrogate or electronic controls may be used to fulfill the daily
QC requirement. In all cases, the appropriate number of controls
(two levels per day for most analytes) must be included and the
results must be documented and reviewed to ensure the adequacy
of the testing process.
In all cases, documentation of both the QC results and the
specific remedial action to "out of control results"
must be available to the inspector.
High Complexity Tests
This category includes those tests that are modified by the
laboratory, developed by a laboratory, or a test classified as
high complexity under CLIA. Laboratories under CLIA and COLA also
must comply with Section §493.1201 (general QC): The laboratory
must establish and follow written QC procedures for monitoring
and evaluating the quality of the analytical testing process of
each method to assure the accuracy and reliability of patient
test results and reports. Section §493.1202 states for each
test of high complexity performed, the laboratory must meet all
applicable standards of this subpart (subpart K). For statistical
QC, laboratories must be in compliance with the following sections:
- §493.1218: Control procedures are performed on a
routine basis to monitor the stability of the method or test
system; control and calibration materials provide a means to
indirectly assess accuracy and precision of patient test results.
Control procedures must be performed as defined in this section
unless otherwise specified in sections §493.1223 through
§493.1285 (these state specific QC requirements for
blood gases, hematology, etc.).
- §493.1218(b) for each method that is developed in-house,
is a modification of the manufacturers test procedure,
or is a method that has not been cleared by the FDA as meeting
the CLIA requirements for general QC (all highly complex methods),
the laboratory must evaluate instrument and reagent stability
and operator variance in determining the number, type, frequency
of testing calibration or control materials and establish criteria
for acceptability used to monitor test performance during a run
of patient specimen(s).
- §493.1218(d)(1) The stated values of an assayed control
material may be used as the target values provided the stated
values correspond to the methodology and instrument employed
by the laboratory and are verified by the laboratory.
- §493.1218(d)(2) Statistical parameters for unassayed
materials must be established over time by the laboratory through
concurrent testing with calibration materials or control materials
having previously determined statistical parameters. (The
statistical parameters, e.g. mean and SD, for each lot number
must be determined through repetitive testing).
- §493.1218(e) Control results must meet the laboratory's
criteria for acceptability prior to reporting patient test results.
Laboratory criteria for acceptability refers to the particular
control limits or control rules chosen by the laboratory.
What if a control fails to meet criteria
for acceptability?
Regardless how a test has been classified, the regulations
require "remedial action and documentation of this activity.
Section 493.1219(b) Remedial actions states that when: "Results
of controls and calibration materials fail to meet the laboratory's
established criteria for acceptability, all patient test results
obtained in the unacceptable test run or since the last acceptable
test run must be evaluated to determine if patient test results
have been adversely affected and the laboratory must take remedial
action necessary to ensure the reporting of accurate and reliable
patient test results.
Joint Commision on Accreditation of Healthcare
Organizations (JCAHO) Requirements for QC
The requirements identified for CLIA88 and COLA also
are applicable to JCAHO. Under JCAHO [6], the goal of QC is to
achieve quality in laboratory testing and produce the best possible
test results and outcomes. Additional JCAHO requirements are found
for the different test classification categories.
JCAHO and Waived Testing
JCAHO recognizes waived tests as defined by CLIA'88, but identifies
additional QC requirements that include:
- Defined QC checks that at least meet the minimum manufacturer's
recommendations (when no QC requirement is specified, the testing
institution must define a policy);
- Maintenance of appropriate QC and test records;
- Proof of training and continued competence of all testing
personnel;
- Proof all testing personnel have access to current written
procedures for QC and remedial actions.
- Maintenance of QC records, including a mechanism to correlate
or link analyst, QC records, instrument and instrument problems,
and individual patient test results.
- For glucose, at a minimum two levels of controls are required
each day a glucose is performed. QC is focused on each meter
used, not each individual performing the test. Therefore, not
all testing personnel need to routinely perform QC, but each
meter must be validated by QC before testing patient samples.
JCAHO and Moderate Complexity Tests
For moderate complexity tests, JCAHO, for the most part, follows
CLIA'88 and mandates the same seven QC requirements as CLIA88
and COLA and also accepts electronic controls for now. However,
for use of electronic controls, JCAHO requires that the laboratory
verify the manufacturers QC claims and run external (usually
liquid) controls periodically to validate that no change occurred
with the testing system.
JCAHO and High Complexity Testing
Testing sites must follow all the CLIA QC requirements for
high complexity tests as well as for any modified moderate complexity
tests (this includes not following the manufacturers directions)
and tests developed in-house.
JCAHO Standards for Quality Control
In addition to the requirements identified above, JCAHO requires
that all testing sites meet the following standards associated
with quality control.
General QC Requirements:
- (QC.1) Each specialty and subspecialty (of testing) has a
documented quality control program.
- (QC.1.3) The laboratorys QC system includes daily surveillance
of results by appropriate personnel.
- (QC.1.4) The laboratory takes remedial action for deficiencies
identified through QC measures or authorized inspections and
documents such actions.
- (QC.1.5) The laboratory ensures that QC results meet its
criteria for acceptability before it reports patient test results.
Clinical Chemistry QC Requirements:
- (QC.6) Using appropriate controls, the laboratory verifies
each procedure in clinical chemistry at least once each day of
use.
- (QC.6.2) Using repetitive testing, the laboratory establishes
control ranges with valid statistical measurements for each procedure
in chemistry.
- (QC.6.3) The laboratory has established and makes available
to its staff acceptable limits for all standard and reference
QC samples, as well as the action to take when results are outside
satisfactory control limits.
- (QC.6.4) The laboratory established test control limits to
provide results with meaningful clinical applications.
Hematology and Coagulation QC Requirements:
- (QC.7.1) The laboratory verifies each procedure and test
parameter against known standards or controls within the range
of clinically significant values each day of use.
- (QC.7.3) The laboratory has established and makes available
to its staff acceptable limits for all standard and reference
QC samples, as well as the action to take when results are outside
satisfactory control limits.
- (QC.7.4) The laboratory has established test control limits
to provide results with meaningful clinical applications.
College of American Pathologists Laboratory
Accreditation Program (CAP-LAP) Requirements for Quality Control
CAP-LAP's philosophy is that all clinical laboratory
tests need to follow the requirements defined for high complexity
testing under CLIA'88 [7]. Requirements for routine analysis of
QC follow the CLIA requirements in terms of number and frequency,
except controls must be included with all tests (even those identified
as waived tests under CLIA'88). For now, CAP only allows "acceptable"
alternative (electronic) QC for point of care testing. The latest
POCT Checklist (Checklist 30, 1997.0) states that certain unit-use
devices/kits may warrant some combination of instrument, procedural,
and/or electronic controls. Question 30.0550 (Checklist 30) states
that POCT sites using alternative QC must have scientifically
acceptable evidence (documented information) that the entire analytical
process is being evaluated correctly. Except for electronic controls,
CAP requires that control specimens be tested in the same manner
as patient samples.
Qualitative tests (even those with built-in controls) need
to be evaluated with both a positive and negative control each
day of use. For all tests, CAP-LAP requires an audit trail that
ties the patients results with the analyst, instrument and
QC. In addition, the QC program should show evidence of documented
review on the next shift, if no supervisor is on site, and at
least weekly review by the technical supervisor and monthly secondary
review by the director or director's designee.
CAP Guidelines for Quality Control
In the general requirements, CAP states in Checklist #1 1997.1
that the overall QC program for the entire laboratory must
be clearly defined and documented. It must include general policies
and delegation of responsibilities. The QC records should be well-organized
with a defined system to permit regular review by appropriate
supervisory personnel.
- 01.3000) Does the QC program clearly define goals for monitoring
analytic performance, procedures, policies, tolerance limits,
corrective action and related information?
Hematology QC Requirements (Checklist 2, 1996.2)
- (02.2005) For numeric data generated by the hematology laboratory,
are Gaussian or other longitudinal process control statistics
(S.D. and C.V.) calculated at least at monthly intervals to define
analytic precision?
- (02.2010) Are tolerance limits (numeric and/or non-numeric)
fully defined and documented for all hematology and coagulation
control procedures?
- (02.2580) Does the laboratory use preserved or stabilized
whole blood preparations for longitudinal process control?
- (02.2582) Does the laboratory use retained, previously analyzed
patient whole blood samples for longitudinal process control?
- (02.2587) If assayed controls are used for CBC instruments,
do control values correspond to the methodology and have target
values been verified by the laboratory for quantitative tests?
- (02.2588) If unassayed controls are used, has a statistically
valid target range been established for each lot by repetitive
analysis in runs that include previously tested control materials?
Automated/General Chemistry QC Requirements (Checklist 3,
1996.2)
- (03.2600) For quantitative tests, are control specimens (with
a matrix appropriate for the method) at more than once concentration
(when available) used for all tests with each analytical run?
- (03.2605) If assayed controls are used, do control values
correspond to the methodology and have target values been verified
by the laboratory for quantitative tests?
- (03.2610) If unassayed controls are used, has a statistically
valid target range been established for each lot by repetitive
analysis in runs that include previously tested control materials?
- (03.2625) Are tolerance limits defined for control procedures?
Note: For tests with numeric results, recovery ranges supplied
by manufacturers of assayed controls must not be substituted
for QC range limits determined the laboratory on its own equipment.
References
- U.S. Department of Health and Human Services. Medicare, Medicaid
and CLIA programs: Regulations implementing the Clinical Laboratory
Improvement Amendments of 1988 (CLIA). Final rule. Fed Regist
1992; 57:7002-186.
- Laessig RH, Ehrmeyer SS: New Poor Mans (Person)
Guide to Meeting the Regulations. R & S Consultants, Madison
WI. 1997.
- Accreditation Manual. Commission on Office Accreditation
(COLA). Columbia, MD, 1996.
- U.S. Department of Health and Human Services. Medicare, Medicaid
and CLIA programs: Regulations implementing the Clinical Laboratory
Improvement Amendments of 1988 (CLIA) and Clinical Laboratory
Act program fee collection. Fed Regist 1993:58:5215-37.
- U.S. Department of Health and Human Services. Medicare, Medicaid
and CLIA Programs: Extension of Certain Effective Dates for Clinical
Laboratory Requirements Under CLIA. Fed Regist 1997;62:25855-58.
- Accreditation Manual for Pathology and Clinical laboratory
Services. Joint Commission on Accreditation of Healthcare Organizations
(JCAHO). Oakbrook Terrace, IL, 1996.
- Laboratory Accreditation Program. College of American
Pathologists (CAP). Northfield, IL, 1997.
Biography: Sharon S. Ehrmeyer, Ph.D.
Sharon Ehrmeyer, Ph.D., MT(ASCP) is Professor of Pathology
and Laboratory Medicine and Director of the Medical Technology
Program at the University of Wisconsin in Madison, Wisconsin.
Dr. Ehrmeyer is active in the American Association for Clinical
Chemistry, the American Society for Clinical Laboratory Science
and the National Committee Clinical Laboratory Standards where
she serves on the Board of Directors and chairs its pH/Blood Gas
Committee. Dr. Ehrmeyer gives numerous presentations on laboratory
regulations (CLIA, JCAHO, CAP and COLA), point of care testing
and various quality issues. Her research interests focus on clinical
laboratory quality and the impact of government regulations on
laboratory practices.
Copyright © 2000. All rights reserved
Westgard QC, 7614 Gray Fox Trail, Madison WI 53717
Call 608-833-4718 or e-mail westgard@westgard.com
A Message from
JOW
QC Lessons | QC
Applications | Questions | Multirule
CLIA Requirements |
What's new? | Catalog
| Demo Download
Home | Glossary
| ARCHIVES | Links
| Feedback