Selection of analytical methods and control procedures can be accomplished with a systematic quality-planning process. Key elements in this planning process are the definition of the quality requirement for the test and the translation of that quality requirement into operating specifications for imprecision, inaccuracy, and quality control (QC). A thyroxine example is provided to illustrate the different formats for defining quality requirement and the use of an OPSpecs chart for setting specifications for the imprecision and inaccuracy that are allowable and for selecting the appropriate control rules and the total number of control measurements for a statistical QC procedure. Quality planning for immunoassays is complicated by the need to consider multiple formats of quality requirements, multiple medical decision levels and the analytical performance at each of those decision levels, and multiple QC designs. Improvements in quality-planning technology should be able to overcome these complications. Quality-planning tools, technology, and training can be delivered via the Internet to support applications in individual laboratories.
"Quality control" (QC) implies that a defined level of quality is assured or guaranteed for the laboratory tests being performed. Unless the quality to be achieved has been defined, the traditional laboratory practice of analyzing control samples actually amounts to "arbitrary control." However, the term "arbitrary" wouldn't be very comforting to the physicians and patients who are the customers and consumers of the test results. Quality control sounds good, even though the actual control practice in most laboratories is arbitrary.
To put quality into quality control, the desired level of quality must be defined in the beginning to guide the selection of analytical methods AND quality control procedures. Then the management objectives are to select methods with appropriate imprecision and inaccuracy and to select statistical control rules and numbers of control measurements to detect medically important errors.
Some guidelines for planning QC procedures for immunoassays are available in the literature [1-3]. The steps of a systematic planning process that utilizes charts of operating specifications (OPSpecs charts) have been presented by Mugan, Carlson, and Westgard, along with seven specific examples for tests performed on an automated analyzer [1]. Seth has illustrated how QC procedures can be selected with the aid of critical-error graphs. Carey has discussed practical aspects of selecting appropriate QC procedures in an essay on Tips for managing the quality of immunoassays that appears on this website (including four example applications).
| A systematic process is needed to carefully consider the many variables and factors that affect the outcome of an analytical testing process. The first step is to define the quality requirement for the diagnostic test of interest. Then, as shown in Figure 1, there are two variations of the planning process, depending on whether the purpose is to select the method of analysis or to select the QC procedure. |
|
| A more detailed outline of the QC selection process is provided in Figure 2, which identifies the following steps: |
|
Applications of this step-by-step process can be found elsewhere on this website.
The two key elements in the planning process are the quality requirement and the OPSpecs chart. The quality requirement guides the selection of the method and the QC procedure. The OPSpecs chart [4,5] is a tool for translating a quality requirement into the imprecision and inaccuracy that are allowable for the method and the control rules and N that are required for the QC procedure, i.e., the operating specifications needed at the bench level to assure the quality of test results.
The most useful format is an interval statement that encompasses the overall or total variation expected in a test result due to the factors of interest in the analytical and pre-analytical parts of the testing process. For the analytical part of a testing process alone, the quality requirement can be stated as an allowable total error, such as often specified by criteria for acceptability in proficiency testing programs [6] or as calculated from biologic goals for imprecision and inaccuracy as recommended by Hyltoft Petersen et. al [7]. For the pre-analytical plus analytical parts, a quality requirement can be stated in the form of a medically important change or decision interval [8]. This clinical requirement requires that pre-analytical factors such as within-subject biological variation of the patient be taken into account when assessing the operating specifications for imprecision, inaccuracy, and QC.
For thyroxine, for example, the US CLIA criterion for acceptability is give as 20% or plus/minus 1.0 mcg/dL, whichever is greater [6]. The European interim biologic goal for imprecision is 4.1% and for inaccuracy is 6.8% which give a calculated allowable total error of 13.6%. Skendzel, Barnett, and Platt [9] recommend that a change of 33% in a patient's test results would be medically significant. Note that the within-subject biologic variation, which is given as 5.8% by Fraser [10], must be taken into account to properly consider the imprecision and inaccuracy allowable for the method itself [11].
| Figure 3 shows the relationships between these different types of quality requirements and the operating specifications that are needed to manage the analytical quality of a laboratory testing process. |
|
The relationship between imprecision, inaccuracy, and QC can be shown graphically by an OPSpecs chart, which is derived from an "error budget" or quality-planning model [12]. An analytical quality-planning model is used to relate the allowable total error to the imprecision, inaccuracy, and sensitivity of the QC procedure [6]. A clinical model relates the decision interval to pre-analytical factors, such as within-subject biologic variation, as well as the analytical factors - imprecision, inaccuracy, and QC [8].
| For example, Figure 4 shows an OPSpecs chart that has been prepared for a TEa of 20% (the CLIA criterion for acceptable performance on proficiency testing surveys), as indicated at top. |
|
To use an OPSpecs chart for setting specifications for the imprecision and inaccuracy of a method, you define the control rules and N desired for routine laboratory practice. For example, if you want to use a 13s control procedure with N=3 in your laboratory, the method needs to have a CV of 4.0% or better if bias is zero (the x-intercept for the operating limits of the 13s rule with N=3). If you are willing to use a multirule procedure with an N of 6, the method can have a CV as high as 6.0% (the x-intercept of the multirule procedure with N of 6).
To use an OPSpecs chart to select an appropriate QC procedure,
you need to know the imprecision and inaccuracy of the method.
Given a method with an observed CV of 5.0% and an observed bias
of 1.0%, these values become the x and y coordinates of the method's
"operating point" as shown in the figure. Any QC procedures
whose operating limits are above the point provide appropriate
QC. The 4-rule multirule procedure with an N of 6 could be selected
here.
OPSpecs charts are very simple to use, as illustrated above. The difficulties are to understand why they work (i.e., the theory) and to prepare the individual charts needed for each individual application (i.e., the calculations and graphic presentation). Obviously, the calculations and the graphics can best be handled by computer, either by an electronic spreadsheet that facilitates manual selection [6,8] or by a specialized PC program that can automate the selection process on the basis of user-defined criteria and logic [13,14].
Even with a computer program, the quality planning for immunoassays poses special difficulties due to multiple formats for quality requirements, multiple decision levels, and multiple estimates of method performance at different decision levels. In addition, because immunoassay measurement procedures are generally not as precise as highly automated chemistry and hematology systems, multiple QC designs for "startup" and "monitoring" may be needed to effectively manage the quality of routine testing processes [1]. Because these "multiple" factors require preparation of additional OPSpecs charts, it would be advantageous to compare their effects side-by-side, all at once, as a single quality-planning exercise, in a single computer file.
| Future computer programs should be able to address these needs for improvements in quality-planning technology. For example, Figure 5 shows the QC Instructions window of a prototype PC program that considers all these "multiple" factors (multiple formats of quality requirements, multiple design levels, multiple estimates, of method performance, and multiple designs). Three different QC designs that can be selected simultaneously and each can consider up to three different types of quality requirements, up to four different medical decision levels, and estimates of method performance at each of these decision levels. |
|
Quality-planning tools and technology will not by themselves improve the management of quality in healthcare laboratories! Training is also needed to support the efforts of the directors, managers, supervisors, and analysts who work in these laboratories. Fortunately, training can now be delivered directly to a laboratory through internet courses, such as the Basic QC Practices course which is offered by the ASCLS. This course can be accessed at the AACC website at the address http://www.ascls.org/wqc/. A training CD for advanced QC planning is also available; a description of the course can also be found on this website which provides an outline of the objectives, materials, and lessons. As part of the course, participants can download a PC computer program for QC planning, along with tutorials for learning operation of the program. Thus, the tools, technology, and training can be delivered directly to laboratories anywhere and anytime through the internet. For laboratories where internet access is slow or limited, this course can be provided on CD-ROM.