Rajavithi Hospital Laboratory Sigma Verification of Performance

Rajavithi Hospital Laboratory joins the Sigma Verification of Performance Program. An impressive debut with 6 instruments.

Sigma Verification of Rajavithi Hospital Laboratory

Verification: February 1st, 2021
Duration of Verification: February 1st, 2021 through March 31st, 2022

 In 2021, Rajavithi Hospital Laboratory joined the Westgard Sigma Verification of Performance Program.

From their application

"Rajavithi Hospital is a 1,200-bed medical center accommodating 40,000 in-patients and 1,000,000 out-patients yearly or a daily average of 4,000 patients, with 200,000 referral patients (both in and out patients). The hospital provides medical service at standard health promotion in tertiary levels and above, including quality referral system. There are more than 200 medical doctors, over 800 professional nurses and specialized staff and over 4,000 other supporting staff who are ready to provide medical services in varied specialties and healthcare dimensions."

Summary of Analyte Performance

Instrument performance in Chemistry were verified.

Chemistry Assay

Verified on Ac02620

Verified on Ac02651

Verified on Ac02652

Albumin

Six Sigma

Four Sigma

Six Sigma

Alk Phos

Six Sigma

Six Sigma

Six Sigma

ALT

Four Sigma

Six Sigma

Four Sigma

AST

Four Sigma

Four Sigma

Six Sigma

Bilirubin, Direct

Four Sigma

Six Sigma

Six Sigma

Bilirubin, Total

Four Sigma

Four Sigma

Four Sigma

Calcium

Six Sigma

Six Sigma

Six Sigma

Chloride

Four Sigma

Five Sigma

-

Cholesterol

Six Sigma

Six Sigma

Six Sigma

Creatinine

Six Sigma

Six Sigma

Six Sigma

Creatinine Kinase (CK)

Six Sigma

-

Six Sigma

Glucose

Six Sigma

Five Sigma

Six Sigma

HDL

Six Sigma

Six Sigma

Six Sigma

Iron

Four Sigma

-

Six Sigma

LDL

Six Sigma

Six Sigma

Six Sigma

Magnesium

Six Sigma

Six Sigma

Six Sigma

Potassium

Six Sigma

Six Sigma

Six Sigma

Phosphorous

Six Sigma

Six Sigma

Six Sigma

Protein, Total

Five Sigma

Five Sigma

Six Sigma

Triglycerides

Six Sigma

Six Sigma

Six Sigma

IUBC

Four Sigma

-

Four Sigma

Uric Acid

Six Sigma

Six Sigma

Six Sigma

Chemistry Assay

Verified on Ac02653

Verified on Ac02654

Verified on Ac02655

Albumin

Six Sigma

Six Sigma

Six Sigma

Alk Phos

Six Sigma

Six Sigma

Five Sigma

ALT

Five Sigma

Six Sigma

Six Sigma

AST

Five Sigma

Six Sigma

Six Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Six Sigma

Bilirubin, Total

Four Sigma

Five Sigma

Five Sigma

Calcium

Four Sigma

Six Sigma

Six Sigma

Chloride

-

Four Sigma

Six Sigma

Cholesterol

Five Sigma

Six Sigma

Six Sigma

Creatinine

Six Sigma

Six Sigma

Six Sigma

Glucose

Six Sigma

Six Sigma

Six Sigma

HDL

Six Sigma

Six Sigma

Six Sigma

LDL

Six Sigma

Six Sigma

Six Sigma

Magnesium

Six Sigma

Six Sigma

Six Sigma

Potassium

Six Sigma

Six Sigma

Six Sigma

Phosphorous

Six Sigma

Six Sigma

Six Sigma

Protein, Total

Five Sigma

Five Sigma

Six Sigma

Sodium

-

Four Sigma

Five Sigma

Triglycerides

Six Sigma

Six Sigma

Six Sigma

Uric Acid

Six Sigma

Six Sigma

Six Sigma

Six Sigma: World Class Quality. Possible to use 3s to 3.5s limits and a minimum of controls

Five Sigma: Excellent Quality. Possible to use 3s limits or a 1:3s/2:2s/R:4s Westgard Rule

Four Sigma: Good Quality. Possible to use 1:3s/2:2s/R:4s/4:1s Westgard Rule

 Visual Summary of Performance of Rajavithi Hospital Laboratory

2021 Rajavithi Sigma VP 1 

2021 Rajavithi Sigma VP 2

2021 Rajavithi Sigma VP 3

2021 Rajavithi Sigma VP 4

2021 Rajavithi Sigma VP 5

2021 Rajavithi Sigma VP 6

 

 Congratulations again to Rajavithi Hospital Laboratory and their staff.

 

 

 

 

Joomla SEF URLs by Artio

About this website

WESTGARD QC promotes the latest news, education, and tools in the quality control field. Our goal is to bring tools, technology and training into today's healthcare industry — by featuring QC lessons, QC case studies and frequent essays from leaders in the quality control area. This is also a reference source for quality requirements, including CLIA requirements for analytical quality. This website features the best explanation of the Multirule ("Westgard Rules") and how to use them. For laboratory and healthcare professionals looking for educational and reference material in the quality control field.