WQC GLOSSARY
A | B | C
| D | E-I | J-M
| N-O | P | Q
| R-S | T-Z
Glossary of Terms Definitions of important terms are obtained
from documents of professional societies and organizations such
as the International Federation of Clinical Chemistry [IFCC],
the American Society for Quality [ASQ], and the National Committee
for Clinical Laboratory Standards [NCCLS], or, in some cases,
are based on the use of these terms in the context of the approach
to analytical quality management that is being recommended in
these lessons.
RE. A term that describes a change in random
error from the stable imprecision of the method (smeas).
A value of 2.0 indicates a doubling of the original method standard
deviation, whereas a value of 1.0 represents the original stable
standard deviation. Used in our quality planning models or error
budget equations to indicate the increase in random error that
can be detected by a control procedure.
REcrit
is a special case that represents the increase in random error
that needs to be detected to maintain a defined quality requirement.
SE. A term that describes a change in systematic
error in multiples of the standard deviation (smeas)
observed for the method under stable conditions. A value of 2.0
would indicate a systematic shift equivalent to two times smeas.
Used in our quality planning models or error budget equations
to indicate the size of systematic error that can be detected
by a control procedure.
SEcrit
is a special case that represents the systematic shift that needs
to be detected to maintain a defined quality requirement.
- Accuracy. "Agreement between
the best estimate of a quantity and its true value. It has no
numerical value. See inaccuracy" [IFCC].
- Allowable total error, TEa. An analytical
quality requirement that sets a limit for both the imprecision
and inaccuracy that are tolerable in a single measurement or
single test result.
- Analyte. The substance to be measured.
- Analytical error. "Difference between the estimated
value of a quantity and its true value. This difference (positive
or negative) may be expressed either in the units in which the
quantity is measured, or as a percentage of the true value"
[IFCC]. Used here to mean the difference between a patient's
test result produced by the analytical process and the true or
correct value for that sample.
- Analytical quality assurance (AQA). Used with charts
of operating specifications (OPSpecs charts) to indicate the
level of assurance for detecting critical sized errors. For example,
90% AQA(SE) means there will be at least a 90% chance of detecting
the critical systematic error when operating within the allowed
limits for imprecision and inaccuracy for the given control rules
and total number of control measurements (N).
- Analytical quality requirement. Used here to refer
to a quality requirement in the form of an allowable total error
(TEa). Often defined on the basis of proficiency testing
criteria for acceptable performance, such as the CLIA requirements
for regulated analytes.
- Analytical run. Defined by CLIA as an interval within
which the accuracy and precision of a testing system is expected
to be stable. Cannot be longer than 24 hours. This definition
is based on that given in NCCLS document C24A that defines a
run as an interval, period of time, or number of specimen, for
which the precision and accuracy of the method is expected to
remain stable; between which events may occur that could cause
errors that are important to detect.
- Assayed control material. A control solution or control
material for which the manufacturer defines the results expected
for different tests and different methods. These "bottle
values" are summarized on assay sheets and are useful for
selecting control materials, however, they should not generally
be used for setting the control limits because they values usually
include a between-laboratory component of variation that makes
the control limits too wide for use in an individual laboratory.
- Bias. A systematic difference between
an observed value and some measure of the truth. Generally used
to describe the inaccuracy of a method relative to a comparative
method in method evaluation or to a comparative group in proficiency
testing.
- Biasmeas. Used here to represent the bias
of a measurement procedure relative to a comparative method or
a comparative group in proficiency testing.
- Clinical quality requirement. Used
here for a quality requirement that states the medically important
change in a test result or describes the gray zone or decision
interval for interpreting a test result.
- Coefficient of variation, CV. The relative standard
deviation, i.e., the standard deviation expressed as a percentage
of the mean [CV=100(s/x)].
- Control chart. "A graphical method for evaluating
whether a process is or is not in a 'state of statistical control.'
The determinations are made through comparison of the values
of some statistical measure(s) for an ordered series of samples,
or subgroups, with control limits" [ASQ]. In healthcare
laboratories, the Levey-Jennings chart is commonly used to plot
the result observed for a stable control material versus time,
usually the day or run number.
- Control limits. "Limits on a control chart which
are used as criteria for signalling the need for action, or for
judging whether a set of data does or does not indicate a 'state
of control'" [ASQ]. Used here to refer to the defined limits
or ranges of results expected due to the random error of the
method, and beyond which some course of action should be taken.
It is common in clinical laboratories to use Levey-Jennings control
charts with limits set as either the mean plus or minus 2 standard
deviations, or the mean plus or minus 3 standard deviations.
- Control material, control product. A control solution
that is available, often commercially, liquid or lyophilized,
and packaged in aliquots that can be prepared and used individually.
- Control measurements, control observations. The analytical
results obtained for control solutions or control materials (that
are analyzed for purposes of quality control).
- Control procedure, QC procedure. The protocol and
materials that are necessary for an analyst to assess whether
the method is working properly and patient test results can be
reported - that part of an analytical process that is concerned
with testing the quality of the analytical results, in contrast
to the measurement procedure, which produces the result. A control
procedure can be described by the number of control measurements
and the decision criteria (control rules) used for judging the
acceptability of the analytical results.
- Control rule. A decision criterion for interpreting
control data and making a judgement on the control status of
an analytical run. Symbolized here by AL, where A
is the abbreviation for a particular statistic or states the
number of control measurements, and L is the control limit. An
analytical run is rejected when the control measurements fulfill
the stated conditions, i.e., when a certain statistic or number
of control measurements exceeds the specified control limits.
- Criteria for acceptability. CLIA's term for the decision
criteria applied to assess the validity of an analytical run.
Another name for QC procedure, with emphasis on definition of
the decision criteria or control rules for interpreting control
measurements.
- Critical-error graph. A power function graph on which
is imposed the critical-error that needs to be detected. Facilitates
the estimation and comparison of the probability for error detection
by different QC procedures.
- Critical random error,
REcrit.
The size of random error that causes a 5% maximum defect rate
for the analytical process. Calculated as (TEa-
biasmeas)/1.65smeas, where TEa
is the allowable total error, biasmeas is the inaccuracy,
and smeas is the imprecision (standard deviation)
of the measurement procedure.
- Critical systematic error,
SEcrit.
The size of the systematic error that needs to be detected to
maintain a defined quality requirement. Calculated as [(TEa
- biasmeas)/smeas] - 1.65, where TEa
is the allowable total error, biasmeas is the inaccuracy,
and smeas is the imprecision (standard deviation)
of the measurement procedure.
- Cumulative control limits. Control limits calculated
from estimates of the mean and standard deviation that represent
a time period longer than a month. Common practice is for laboratories
to calculate monthly control statistics. Cumulative statistics
can be easily calculated from monthly statistics that tabulate
the sum of the individual values and the sum of the squares of
those values. See the lesson QC - The
Calculations for more detailed discussion, equations, and
examples.
- Decision Interval, Dint.
Used here to represent the gray zone, or interval of uncertainty,
in interpreting a test result. An example is the NCEP (National
Cholesterol Education Program) guidelines that indicate a cholesterol
of less than or equal to 200 mg/dL is okay and that a value of
more than or equal to 240 should have followup testing, which
defines a gray zone of 40 mg/dL between 200 to 240, which is
20% at a decision level of 200 mg/dL. Dint can also
be defined more generally by the change in a test result that
is judged to be medically significant.
- Decision level for critical interpretation (Xc).
See medical decision level.
- Distribution. Refers to the spread and shape of a
frequency curve of some variable. A histogram is one way to graphically
display the distribution of test results by showing the frequency
of observations on the y-axis versus the magnitude on the x-axis.
The normal or gaussian curve is one form of a distribution.
- Error detection. See probability
for error detection.
- False rejection. See probability for false rejection.
- Frequency of errors, frequency of occurrence of analytical
errors, f. A performance characteristic of a measurement
procedure that describes how frequently analytical errors are
expected to occur. Related to the stability of the measurement
procedure.
- Gaussian curve. Gaussian distribution. Normal curve. Normal
distribution. Refers to a symmetrical bell-shaped distribution
whose shape is given by a specific equation (called the normal
equation) in which the mean and standard deviation are variables.
It is commonly assumed that the random error of an analytical
method fits the Gaussian distribution and therefore can be characterized
by calculating the standard deviation. The standard deviation
is not a valid statistic if a distribution is not Gaussian.
- High complexity tests. A CLIA category of tests that
has the most demanding QC requirements. Includes any tests developed
by the laboratory, modified by the laboratory, or manufacturer's
tests that have been classified as high complexity.
- Imprecision. "Standard deviation or coefficient
of variation of the results in a set of replicate measurements.
The mean value and number of replicates must be stated, and the
design used must be described in such a way that other workers
can repeat it. This is particularly important whenever a specific
term is used to denote a particular type of imprecsion, such
as between-laboratory, within-day, or between day" [IFCC].
We use the term smeas to represent a measurement procedure's
standard deviation and express it as a percent, or coefficient
of variation, on OPSpecs charts.
- Inaccuracy. "Numerical difference between the
mean of a set of replicate measurements and the true value. This
difference (positive or negative) may be expressed in the units
in which the quantity is measured, or as a percentage of the
true value" [IFCC]. We use the term biasmeas to describe
the average systematic difference between a measurement procedure
and a comparative method and express it in percent on OPSpecs
charts..
- Inherent imprecision, inherent random error. The standard
deviation or coefficient of variation of the results in a set
of replicate measurements obtained when the measurement procedure
is operating under stable conditions.
- Levey-Jennings control chart. A
commonly used control chart in which individual control measurements
are plotted directly on a control chart with limit lines drawn
either as mean ± 2s or mean ± 3s. Time is displayed
on the x-axis usually in terms of days or runs.
- Matrix. Refers to the physical and chemical nature
of the specimen, the substances present, and their concentrations.
The matrix of a control material is an important consideration
in selecting and implementing a QC procedure. In this context,
the matrix refers to the substances and base from which the control
material is prepared, in addition to all the additives such as
spiking materials, preservatives, etc., necessary to make the
product useful.
- Mean. The arithmetic average of a set of values. A
measure of central tendency of the distribution of a set of replicate
results. Often abbreviated by an x with a bar over it.
- Medical decision level, decision level, Xc.
A concentration of analyte where medical interpretation is critical
for patient care. There may be several different medical decision
levels for a particular analyte. Xc should provide
guidance for selecting relevant estimates of stable imprecision,
stable inaccuracy, and matrix inaccuracy. This is analogous to
identifying a critical Target Value (TV) for assessing test performance
and validating QC design.
- Medically important errors. Used here to indicate
errors that, when added to the inherent imprecision and inaccuracy
of a measurement procedure, cause the quality requirement to
be exceeded. Medically important random errors are those increases
in the standard deviation of the measurement procedure that cause
the error distribution to exceed the quality requirement (see
critical random errors). Medically important systematic errors
are those shifts in the mean of the error distribution that cause
the error distribution to exceed the quality requirement (see
critical systematic error).
- Method validation. The process of testing a measurement
procedure to assess its performance and to validate that performance
is acceptable. The magnitudes of the analytical errors are experimentally
determined and their acceptability for the application of the
method is judged versus defined requirements for quality in the
form of maximum allowable errors.
- Model. A mathematical equation that describes the
behavior of a process as a function of its important characteristics.
- Moderate complexity tests. A CLIA category of tests
that includes about 75% of all tests performed by healthcare
laboratories, including most automated analytical systems. This
category has more stringent QC requirements than for "waived
tests" or "provider performed microscopy."
- Multi-rule quality-control procedure. A control procedure
that uses two or more control rules for testing control measurements
and determining control status. At least one rule is chosen for
its ability to detect random errors and one to detect systematic
errors.
- Number of control measurements, N.
Used here to indicate the total number of control measurements
available for use in assessing the quality of an analytical run.
We consider N to be the total number of control measurements
available for inspection when using common Levey-Jennings type
QC charts or multirule type QC procedures where it is possible
to combine the measurements from different materials to accumulate
a higher N (and higher error detection) for evaluating control
status. These measurements may be replicates on one level or
material, individual measurements on two or more materials, or
replicate measurements on two or more materials. For example,
if you assay a single material and make two measurements on that
material, N is 2. If you assay two materials (as required by
US CLIA regulations) and make single measurements on each, N
is 2. If you assay two materials and make duplicate measurements
on each, N is 4. If you assay three materials and make single
measurements on each, N is 3. If you assay three materials and
make duplicate measurements on each, N is 6. With the use of
mean/range or cusum type of QC procedures where it is more difficult
to combine the measurements from different control materials,
N is more likely to be the number of replicates on an individual
material.
- Operating specifications (OPSpecs). Used here to describe
the imprecision and inaccuracy that are allowable and the QC
that is necessary to assure, at a stated level, that a defined
quality requirement will be achieved in routine operation.
- OPSpecs chart. A plot of the inaccuracy (on the y-axis)
and the imprecision (on the x-axis) that are allowable for different
QC procedures. The chart is prepared for a defined quality requirement
and for a stated level of analytical quality assurance (AQA).
Readily available in workbook format in the OPSpecs Manual and
also easily prepared by the QC Validator computer program.
- Performance characteristics. Those
properties that describe how well a procedure performs. For a
control procedure, the performance characteristics are the probabilities
for error detection and false rejection, or the average run lengths
for rejectable and acceptable quality. For a measurement procedure,
the performance characteristics include analytical range, precision,
accuracy, interference, recovery, and also the frequency and
duration of analytical errors.
- Power curve. A line on a power function graph that
describes the performance of a certain control rule and N.
- Power-function graph. A graphical presentation of
the performance characteristics of QC procedures that describes
the probability for rejection (on the y-axis) versus the size
of analytical error occurring (on the x-axis) for stated control
rules and numbers of control measurements.
- Precision. "The agreement between replicate measurements.
It has no numerical value" [IFCC, NCCLS]. See also Imprecision.
- Preventive maintenance. The renewing or refurbishing
or critical parts of a method on a regular basis to prevent malfunctions.
- Primary standard material. "Substance of known
chemical composition and sufficient purity to be used in preparing
a primary standard solution" [IFCC].
- Primary standard solution. "Solution used as
calibration standard in which the concentration is determined
solely by dissolving a weighed amount of primary standard material
in an appropriate solvent, and making a stated volume or weight"
[IFCC].
- Probability, p. The likelihood an event will occur,
usually stated as a decimal fraction between 0 and 1, 0 meaning
that the event will never occur and 1 meaning that the event
will always occur. For example, p=0.05 means there is a 5% chance
that an event will occur. Commonly used in quality control to
describe the chance that a run will be rejected.
- Probability for error detection, Ped. A
performance characteristic of a QC procedure that describes how
often an analytical run will be rejected when results contain
errors in addition to the inherent imprecision of the measurement
procedure. Ideally, Ped should be 1.00 for errors
that are medically significant. In practice, we generally aim
for a Ped of 0.90 when selecting and designing QC
procedures.
- Probability for false rejection, Pfr. A
performance characteristic of a QC procedure that describes how
often an analytical run will be rejected when there are no errors
occurring, except for the inherent imprecision of the measurement
procedure. Ideally, Pfr should be 0.00. In practice,
we generally aim for a Pfr of 0.05 or less.
- Process capability. An industrial term used to describe
how the inherent variability of a production process under stable
operation compared to the allowable variation. SEcrit
is an index of process capability for an analytical testing process.
- Process stability. Used here to characterize the performance
of the measurement procedure in terms of the frequency of analytic
runs having medically important errors (f) that invalidate the
medical usefulness.
- Proficiency testing, PT. Used in the US to describe
a program of external quality assessment whereby specimens are
submitted to laboratories for their analysis with the purpose
of grading the performance of the laboratory.
- Proficiency testing criteria for acceptable performance,
PT criteria. Defined limits about a target value (TV) that
are used to classify analytical performance as acceptable or
not. CLIA defines PT criteria for about 80 regulated analytes,
using a format of TV ± a stated %, TV ± a fixed
concentration, or TV ± 3 SD, where the SD is usually a
group standard deviation from a PT survey. These PT criteria
should be interpreted as total error criteria because only single
measurements can be made on PT specimens and the test result
is subject to both random and systematic errors.
- Provider performed microscopy, PPM. A special subset
of moderately complex tests that may be performed by physicians,
dentists, nurse practioners and midwives, and physician assistants
as part of a patient examination.
- QC acceptability criteria. The term
used by CLIA to indicate the decision criteria or control rules
used to monitor test performance during a run of patient specimens.
- QC planning process, quality planning process. The
steps to be followed to select control rules, N, and a Total
QC strategy that are appropriate for the quality needed and the
imprecision and inaccuracy observed for a laboratory test.
- Quality planning model. Term used to describe an equation
that shows the additive effects of different factors that influence
the variation of a test result. The analytical model relates
the imprecision and inaccuracy of the measurement procedure and
the sensitivity of the control procedure to the total analytical
error that is allowable. The clinical model includes the analytical
components plus pre-analytical components, such as within-subject
biological variation, etc., and relates them to the clinical
decision interval or gray zone for interpreting a test result.
- Random error, RE. An error that
can be either positive or negative, the direction and exact magnitude
of which cannot be exactly predicted. In contrast, systematic
errors are always in one direction.
- Run. See analytical run.
- Sample. "The appropriate representative part
of a specimen which is used in the analysis." [IFCC]
- Specimen. "Material available for analysis."
[IFCC]
- Standard. "Material or solution with which the
sample is compared in order to determine the concentration or
other quantity. The compound term calibration standard should
be used whenever needed to avoid confusion with other technical
or colloquial meanings of the word standard." [IFCC]
- Standard deviation, s. A statistic that describes
the dispersion or spread of a set of measurements about the mean
value of a gaussian or normal distribution. Calculated from the
equation:
where n is the number of measurements, and xi is an
individual measurement.
- Standard deviation index, SDI. Generally used in reports
from proficiency testing (PT) survey to describe how far a PT
result is from the target value (TV). An SDI of +2.0 means that
the laboratory's reported result is 2 standard deviations of
the group above the target value or mean of the group.
- Statistical quality control. Those aspects of quality
control in which statistics are applied, in contrast to the broader
scope of quality control which includes many other procedures,
such as preventive maintenance, instrument function checks, and
performance validation tests. Statistical QC procedures are often
used to monitor routine performance of a method and to alert
the laboratory when the performance of a method changes.
- Statistical control limits. As used with Levey-Jennings
and Westgard multirule types of QC procedures, these are the
lines drawn on control charts to define the range of results
expected due to the random error of the method. The limits are
often obtained from a group of 20 or more measurements on a particular
control material by calculating the mean and standard deviation,
then using multiples such as the mean plus/minus 3s, 2s, or 1s
to establish rejection limits for different control rules.
- Statistical process control. A general term used to
describe those aspects of a control system in which statistics
are applied to determine whether observed measurements fall within
the range expected due to the random variation of the process.
Industrial process control procedures provided the basis for
introduction of statistical control in healthcare laboratories,
however, industrial process control procedures often use the
mean and range of a group of control measurements (e.g., Shewhart
mean and range charts), whereas healthcare applications tend
towards the use individual measurements or individual-value control
charts, such as the Levey-Jennings chart.
- Systematic error, SE. An error that is always in one
direction and is predictable, in contrast to random errors that
may be either positive or negative and whose direction cannot
be predicted.
- Test complexity. Refers to the CLIA
system of classifying tests into categories on the basis of the
difficulty of measurement and interpretation. Categories include
waived, provider performed microscopy, moderate complexity, and
high complexity.
- Total error, TE. The net or combined effect of random
and systematic errors.
- Total error requirement. See allowable total error.
- Total imprecision. The random error observable over
a period of many runs and many days.
- Total QC strategy. The balance of the efforts expended
on statistical QC, preventive maintenance, instrument function
checks, method performance tests, and quality improvement.
- True value. Generally used to indicate that this is
the correct analytical concentration or result.
- Variable. A quantity of interest, whose value or magnitude
fluctuates or changes.
- Variance. The standard deviation squared. If there
are independent sources of errors, the variance of the total
error is the sum of the variances due to the individual sources
of error.
- Waived tests. A specific category of tests defined
by CLIA, such as dipstick tests, fecal occult blood, urine pregnancy
tests, erythrocyte sedimentation rates, blood glucose monitoring
devices, etc., which are subject to the lowest level of regulation.
The main requirement for QC is to follow the manufacturer's directions.
- Westgard rules, Westgard multi-rule control procedure.
A control procedure that uses a series of control rules to test
the control measurements. A 12s rule is used as a
warning, followed by use of 13s, 22s, R4s,
41s, and 10x as rejection rules.
- Within-run imprecision. The random error observable
within the time period of a single analytical run.
- z-score, z-value. A calculated number that tells how
many standard deviations a control result is from its mean value,
e.g., a control result of 112 on a material having a mean of
100 and a standard deviation of 5 has a z-score of +2.4, i.e.,
it is 2.4 standard deviations above its mean.
Copyright © 2000. All rights reserved
Westgard QC, 7614 Gray Fox Trail, Madison WI 53717
Call 608-833-4718 or e-mail westgard@westgard.com
A Message
from JOW
QC Lessons | QC Applications | Questions
| Multirule
CLIA Requirements
| What's new? | Catalog
| Demo Download
Home | Glossary
| ARCHIVES | Links
| Feedback