BIOLOGICAL VARIABILITY DATA BANK,
PART II: ANALYTES L-Z


María Ángeles Sebastián-Gámbaro,
Francisco Javier Lirón-Hernández,
Xavier Fuentes-Arderiu
Servei de Bioquímica Clínica
Ciutat Sanitària i Universitària
de Bellvitge L'Hospitalet de Llobregat,
Barcelona, Spain
Correspondence:
Dr. X. Fuentes-Arderiu
Servei de Bioquímica Clínica
Ciutat Sanitària i Universitària de Bellvitge
08907 L'Hospitalet de Llobregat Barcelona, Spain

Part I, Analytes A-L
References


Table Description. Medians of the coefficients of inter-individual (med CVBb ) and intra-individual (med CVBw ) biological variation of biological quantities in healthy subjects estimated in different studies with different time spans. [The symbols used in this table are: h = hour; d = day; wk = week; mo = month; a = year; r = mass concentration; b = catalytic concentration ; m = mass; c = substance concentration; n/N= entitic amount of substance; n = amount of substance; p = partial pressure; arb = arbitrary; t = time; C = number concentration; f = volume fraction; V/N= entitic volume; f = volume; x = substance fraction; l = length; z = catalitic activity; NO = no data available]. According to ISO, IUPAC and IFCC recommendations the comma is used as decimal sign.

Quantity

time span
range

med
CVBw

med
CVBb

References
Blood-Leukocytes; C 1 mo-1 a 11,2 19,7 19, 53, 82, 95, 96
Serum-Lipoprotein(a); r 5 wk-1 a 10,8 85,8 46, 51, 97, 98
Serum-Lutropin; c arb 1 d-1 a 24,0 29,6 32, 84, 85, 86
Blood-Lymphocytes; C 1mo-1 a 12,3 26,0 53, 82, 95
Leukocytes-Lymphocytes;d 20 wk-1 a 10,6 18,7 53, 82, 96
Serum-a2-Macroglobulin; r 48 h-14 wk 3,3 20,7 7, 17, 22, 27
Erythrocytes-Magnesium(II); c 20 wk- 2 a 5,6 11,2 99, 100
Leukocytes-Magnesium(II); c 20 wk 18,5 12,4 99
Serum-Magnesium(II); c 5 d-2 a 3,2 5,9 16, 19, 24, 25, 33,
37, 56, 79, 100, 101
Urine(daily)-Magnesium(II); c 8 wk-112 d 45,4 37,1 101, 102
Urine(daily)-Magnesium(II); n 8 wk-112 d 38,4 37,6 101, 102
Urine(morning)-Magnesium(II); c 8 wk-112 d 47,2 37,0 101, 102
Urine(random)-Magnesium(II); c 8 wk 57,4 33,0 102
Urine (daily)-Magnesium (II); c
Urine(daily)-Creatininium; c
112 d 35,0 24,0 101
Urine (morning)-Magnesium (II); c
Urine(morning)-Creatininium; c
112 d 35,0 34,0 101

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-b2-Microglobulin; r 21 d-20 wk 4,4 15,5 8, 9
Urine(morning)-a1-Microglobulin; r 2 wk 33,0 58,0 31
Urine(morning)-a2-Microglobulin; r 2 wk 32,0 46,0 31
Blood-Monocytes; C 20 wk 11,7 25,4 82
Leukocytes-Monocytes; d 20 wk 9,8 13,6 82
Serum-Mucinous carcinoma antigen; r 40 wk 5,2 39,3 103
Blood-Neutrophils; C 1 mo-1 a 23,6 32,8 53, 95
Leukocytes-Neutrophils; d 6 mo-1 a 7,4 11,1 53, 96
Serum-5'Nucleotidase; b 48 wk- 1 a 22,3 8,9 13, 74
Serum-Orosomucoid; r 1,4 -24 wk 11,1 30,7 17, 27
Serum-Osteocalcin; c 1 mo 7,3 25,7 12, 34
Urine(daily)-Oxalate; c 8 wk 44,0 18,0 104
Urine(daily)-Oxalate; n 8 wk 42,5 19,9 104
Urine(morning)-Oxalate; c 8 wk 47,5 15,3 104
Urine(random)-Oxalate; c 8 wk 43,1 0,0 104

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-Peptidyl-dipeptidase A; b 10 mo 12,5 27,7 105
Serum-Phosphate (non esterified); c 6 h- 1 a 7,6 11,2 10, 11, 13, 15, 16, 18, 19, 21, 24, 25, 26, 28, 37, 56
Urine(daily)--Phosphate (non esterified); n 5 d-40 wk 18,2 19,7 43, 57, 58
Urine(morning)-Phosphate (non sterified); c 10 wk 43,0 33,9 43
Urine(random)-Phosphate (non sterified); c 2 d-10 wk 45,4 29,8 43, 57
Serum-Phospholipid; c 13 mo 6,9 11,1 106
Blood-Plasma; pH 14 wk 3,5 2,0 61
Blood-Platelets; C 20 wk 9 23,3 82
Erythrocytes-Potassium ion; c 20 wk 0,0 6,5 99
Leukocytes-Potassium ion; c 20 wk 13,6 13,4 99
Serum-Potassium ion; c 6 h-1 a 4,6 4,7 10, 11, 13, 15, 18, 20, 23, 24, 25, 26, 28, 56
Urine(daily)-Potassium ion; c 5 d- 40 wk 28,6 23,2 43, 57, 58
Urine(daily)-Potassium ion; n 5 d- 40 wk 24,4 21,9 39, 43, 57, 58
Urine(random)-Potassium ion; c 2 d-10 wk 38,8 32,4 43, 57

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-Prealbumin; c 21 d 4,4 NO 9
Serum-Progesterone; c 6 mo 31,3 NO 84
Serum-Prolactin; r 1 wk 23,7 52,1 32
Serum-Properdin; r 4 wk 9,5 11,2 7
Serum-Protein; r 6 h- 1 a 2,6 4,8 7, 10, 11, 13, 15, 16, 18, 19, 20, 21, 24, 25, 26, 27, 28, 37, 56
Urine(daily)-Protein; m 5 d- 20 wk 36,1 34,5 43, 57
Urine(daily)-Protein; r 5 d- 20 wk 39,4 17,8 43, 57
Urine(morning)-Protein; r 10 wk 48,4 38,1 43
Urine(random)-Protein; r 2 d-10 wk 48,7 34,9 43, 57
Urine (random)-Pyridinoline; c
Urine(random)-Creatininium; c
4 wk 8,9 14,2 34
Serum-Pyruvate; c 5 d 15,2 13,0 93

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-Retinol; c 1 d-1 mo 20,5 NO 21
Serum-Rheumatoid factors; c arb 6 mo 11,4 24,2 54
Serum-Rubeola virus antibody; c arb 9 mo 6,0 NO 77
Blood-Sedimentation reaction; l 1 a 29,3 NO 107
Pactient-Serum; osmolality 40 wk 1,9 1,4 108
Serum-Sex hormones binding globulin; c 1 d- 1 a 8,7 42,7 85, 86
Erythrocytes-Sodium ion; c 20 wk 1,8 12,4 99
Leukocytes-Sodium ion; c 20 wk 51,0 36,4 99
Serum-Sodium ion; c 6 h- 1 a 0,6 0,6 10, 11, 13, 15, 16, 18, 20, 23, 24, 25, 26, 28, 56
Urine(daily)-Sodium ion; c 5 d- 40 wk 24,0 26,8 43, 57, 58
Urine(daily)-Sodium ion; n 5 d- 40 wk 28,7 18,6 39, 43, 57, 58
Urine(morning)-Sodium ion; c 10 wk 35,8 27,5 43
Urine(random)-Sodium ion; c 2 d- 10 wk 31,1 22,0 43, 57

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-Superoxide dismutase; b 5 d- 5 mo 16,3 0,0 33
Serum-Testosterone; c 1 d- 1 a 9,6 21,3 85, 86
Saliva-Testosterone; c 1 d- 1 a 27,1 28,8 85, 86
Serum-Testosterone; c
Serum-Sex hormone binding globulin; c
1 d- 1 a 12,0 51,6 85, 86
Serum-Thyroglobulin; r 7 d 4,4 12,6 32
Serum-Thyrotropin; c arb 5d- 1a 20,0 29,4 13, 32, 109, 110
Serum-Thyroxine; c 24 h- 1 a 6,0 11,9 13, 15, 17, 32, 37, 109, 110, 111, 112
Serum-Thyroxine(non protein bound); c 5d- 5 wk 7,6 12,2 109, 110
Serum-Tissue polypeptide antigen; c arb 2 a 28,3 133,0 60
Serum-a-Tocopherol; c 1 d- 1 a 15,2 19,9 21, 48
Serum-Toxoplasma gondii antibody; c arb 9 mo 4,0 NO 113
Serum-Transferrin; c 24 h- 24 wk 2,8 2,1 7, 15, 17, 19, 33
Serum-Triacylglycerol lipase; b 40 wk- 1 a 32,2 36,0 13, 40, 42
Serum-Triglyceride; c 13 h- 60 wk 22,0 46,4 13, 15, 16, 17, 18, 19, 20, 21, 27, 37, 44, 45, 46, 48, 49, 50, 51, 62, 63, 64, 65, 66, 67, 68, 69
Serum-Triiodothyronine; c 24 h- 1 a 7,8 17,2 13, 15, 32, 109, 110, 111, 112
Serum-Triiodothyronine(non protein bound); c 5 wk 7,9 22,5 109

Back to Top
Data Bank Part I
References


Quantity

time span
range

med
CVBw

med
CVBb

References
Serum-Urate; c 6 h- 1 a 7,3 18,8 10, 11, 13, 15, 16, 18, 19, 20, 21, 24, 25, 35, 37, 56
Urine(daily)-Urate; c 5 d- 20 wk 25,1 17,4 43, 57
Urine(daily)-Urate; n 5 d- 20 wk 16,8 14,4 43, 57
Urine(random)-Urate; c 2 d 31,3 26,1 43
Serum-Urea; c 6 h- 1 a 11,6 17,4 10, 11, 13, 15, 16, 18, 19, 20, 24, 25, 26, 28, 35, 37, 56
Urine(daily)-Urea; c 5 d- 40 wk 23,2 25,9 43, 57, 58
Urine(daily)-Urea; n 5 d- 40 wk 17,1 21,5 39, 43, 57, 58
Urine(morning)-Urea; c 10 wk 25,0 19,9 43
Urine(random)-Urea; c 2 d- 10 wk 25,3 28,1 43, 57
Serum-VLDL-Cholesterol; c 4 wk 22,5 NO 63
Serum-Zinc(II); c 5 d- 5 mo 6,5 6,1 33, 79

Back to Top
Data Bank, Part I
References
CLIA Quality Requirements
Clinical Quality Requirements.

European Calculated Biologic Allowable Total Errors


Copyright © 2000. All rights reserved.
Westgard QC, 7614 Gray Fox Trail, Madison WI 53717
Call 608-833-4718 or e-mail westgard@westgard.com

A Message from JOW
QC Lessons | QC Applications | Questions | Multirule
CLIA Requirements | What's New? | Catalog | Demo Download
Home | Glossary | ARCHIVES | Links | Feedback