QP-10: Automated Chemistry Applications

Quality planning applications for clinical chemistry tests are perhaps the easiest to get started with because criteria for acceptable performance have been defined by CLIA for many of the routine chemistry tests and estimates of method imprecision and inaccuracy are generally available in the laboratory. It is also helpful that there are documented applications available in the literature. Finally, the high level of automation in chemistry analyzers has improved the performance of chemistry tests, particularly imprecision, thus careful selection of QC procedures can often reduce operating costs! Saving time and money should provide strong motivation for initiating your quality planning applications in the clinical chemistry area.

Quality requirements

CLIA defines criteria for acceptable performance for 27 different tests as part of the proficiency testing guidelines. These criteria are in the form of an allowable total error and are presented as a statement of a target value plus and minus a certain amount. This amount is given in three different forms:

For some tests, it is necessary to define the medical decision level of interest to interpret the quality requirement, e.g., for glucose, acceptable performance is stated to be target value +/- 6 mg/dL or 10%, whichever is greater.

Method imprecision and inaccuracy

Most clinical chemistry methods are classified as moderately complex via CLIA-88 guidelines, which require that the imprecision and inaccuracy of the method be validated through replication and comparison of methods experiments, respectively. Therefore, initial estimates of method imprecision and inaccuracy should be available in all laboratories. Ongoing estimates of method imprecision can be obtained from monthly or cumulative QC data and periodic estimates of method inaccuracy can be made from the difference between the laboratory mean and the group mean in inter-laboratory peer comparison programs, external quality assessment programs, and/or proficiency testing surveys.

Example applications

Some initial examples were presented in lesson 8 to illustrate the selection of QC procedures for glucose, cholesterol and calcium. Real-world applications for these tests have been described in the literature for a multitest chemistry analyzer [1] , where the selection of QC procedures gave the following results:

This study was published in 1990 and therefore predates the 1992 publication of the CLIA-88 criteria for acceptable performance in proficiency testing surveys. Thus, the quality requirements were not exactly the same as those defined by CLIA. Note also that this study makes use of a different quality-planning tool - the critical-error graph - which preceded the OPSpecs chart that was introduced in 1991 [2,3].

Related to this study, we also documented the cost-savings that can be expected from optimized, individualized QC designs [4]. We had initially used a multirule QC procedure for all the tests on this analyzer because that's what was accepted practice on the earlier generation analyzer that was being replaced. Old QC practices are often carried over to new generation analyzers because we assume the new system will have similar performance characteristics. However, improvements in precision and stability may permit more cost-effective QC procedures to be implemented, as demonstrated in this application. The initial changes in QC procedures provided a cost-savings of about $1,450 per month or $17,400 per year, which provides a total savings of $87,000 over the expected 5-year lifetime of the instrument. Later changes in the length of the analytical run increased the savings to $3,000 per month or $36,000 per year, which provides a total savings of $180,000 over the 5-year lifetime of the instrument.

Cholesterol.

The method has an observed standard deviation of 2.7 mg/dL or 1.35% at a decision level of 200 mg/dL. Bias is assumed to be 0.0% because method validation studies show small systematic differences between comparative systems within the laboratory and ongoing comparison studies are used to maintain near-zero biases. Two control materials are to be analyzed each run to fulfill the CLIA requirement. TEa is defined as 20 mg/dL which is 10% at a medical decision level of 200 mg/dL (note the units used in the paper are actually mg/L rather than mg/dL, therefore the values are all higher by a factor of 10).

Glucose.

The observed standard deviation is 1.2 mg/dL or 1.1% at a decision level of 110 mg/dL. Bias is assumed to be zero. TEa was defined as 8.0 mg/dL, which would be 7.2% at a decision level of 110 mg/dL. This is somewhat more demanding than the CLIA requirement of 6 mg/dL or 10% (whichever is greater), which would be 10% at a decision level of 110 mg/dL. Two control materials are to be analyzed.

Chloride.

The method SD was 1.04 mmol/L which is 1.04% at a decision level of 100 mmol/L. Bias is assumed to be zero. The CLIA requirement of 4.0 mmol/L corresponds to 4.0% at a decision level of 100 mmol/L. Two control materials are to be analyzed.

Calcium.

The observed SD was 0.168 mg/dL, which is 1.68% at a decision level of 10.0 mg/dL. TEa was defined as 0.5 mg/dL, which would be 5% at a medical decision level of 10.0 mg/dL (which is much more demanding than the CLIA requirement of 1.0 mg/dL or 10% at a decision level of 10.0 mg/dL).

Planning and implementation strategies

There are lessons to be learned from quality-planning applications in all areas of the laboratory. Routine chemistry applications may be simpler than others, but are still a source of important insights and advice. You're more likely to solve complicated quality-planning problems if you have a good grasp of the solutions to simpler applications.

References

  1. Koch DD, Oryall JJ, Quam EF, Feldbruegge DH, Dowd DE, Barry PL, Westgard JO. Selection of medically useful quality-control procedures for individual tests done in a multitest analytical system. Clin Chem 1990;36:230-233. [also available as PDF file at http://www.westgard.com/multchem.htm]
  2. Westgard JO, Hyltoft Petersen P, Wiebe DA. Laboratory process specifications for assuring the quality in the US National Cholesterol Education Program. Clin Chem 1991;37:656-661.
  3. Westgard JO, Wiebe DA. Cholesterol operational specifications for assuring quality required by CLIA proficiency testing. Clin Chem 1991;37:1938-1944.
  4. Westgard JO, Oryall JJ, Koch DD. Predicting effects of QC practices on the cost-effective operation of a multitest analytical system. Clin Chem 1990;36:1760-1764. [also available as PDF file at http://www.westgard.com/multchem.htm]