QP-11: Blood Gas Applications

Blood gas measurements are often performed in Point-Of-Care (POC) settings as well as larger clinical or pulmonary laboratories. Patients may be monitored during surgery and in the recovery room via the POC tests. Later, when the patients are moved back to their hospital rooms, they may be monitored by measurements performed in a central clinical laboratory or pulmonary function laboratory. The comparability and consistency of test results will be important.

When a new POC analyzer is obtained, the performance of the analyzer should be evaluated by comparison to the measurements being performed in an established laboratory, such as the central clinical chemistry laboratory or the pulmonary function laboratory. Method validation studies will provide estimates of imprecision and bias that can then be used for planning QC procedures.

Quality requirements

The US CLIA criteria for acceptable performance in proficiency testing are as follows:

Blood gas pH Target Value +/- 0.04 pH unit
Blood gas pCO2 Target Value +/- 5 mm Hg or +/- 8% (greater)
Blood gas pO2 Target Value +/- 3 SD

To quantify the requirement for pO2, results from an AACC/CAP proficiency testing survey during 1999 showed the following SDs and CVs for pO2 for "all" analyzers, which represent approximately 3500 laboratories that participated in the survey:

The SD and CV for the 4th specimen (59 mm Hg) appear to be inconsistent with the rest of the figures and further inspection showed that different manufacturers' systems provided large differences in their mean values, demonstrating that the "all" or overall SDs and CVs may reflect a large contribution from systematic errors between different types of analyzers. Averaging the results from the other four specimens gives an SD of 7.1 mm Hg and a CV of 5.5%. The CLIA requirement for pO2 can therefore be estimated to be about +/- 16.5%.

Method imprecision and inaccuracy

To estimate the imprecision of the analyzer, initial method validation studies should include a replication study performed over a period of 20 days. Typically, three different control materials would be analyzed and the SDs and CVs calculated for each level. To estimate inaccuracy or bias, a comparison of methods experiment should be performed on a minimum of 20 and preferably 40 fresh patient samples. The paired data could be analyzed by regression and t-test statistics.

Example applications

For the examples here, the estimates of method CVs are similar to the performance documented in a recent published evaluation study of a POC analyzer [1]. Three different levels of control materials are generally provided for blood gas measurements. Typical levels would be 7.20, 7.40, and 7.60 for pH; 60, 40, and 20 mm Hg for pCO2, and 60, 100, and 140 mm for pO2.

NOTE:

pH

The average method SD for the three control materials is 0.005 pH unit and the method bias is 0.01 pH unit. The CLIA requirement is 0.04 pH units.

pH example application

pCO2

The average CV for the three control materials 2.5%. Method bias is observed to be 0.2 mm Hg at the mean of the comparison data (at approximately 40 mm Hg). The quality requirement at 40 mm Hg would be 5 mm Hg or 12.5%.

pC)2 example application

pO2

The average CV is observed to be 4.1% and the bias between methods is nearly zero. As shown earlier, the CLIA quality requirement is 16.5%, as calculated from results of a proficiency testing survey where the group CV is estimated as 5.5% on the basis of 4 different PT specimens.

pO2 example application

Planning and implementation strategies

It is not unusual to see a range of performance for different blood gas parameters - from excellent for pH, good for pCO2, and marginal for pO2. The method CVs used in the examples here are representative of the performance that can be expected from current portable blood gas analyzers [1], thus there is a need for well designed QC procedures to assure comparability of results between point-of-care and central laboratory applications. The problem of how to QC pO2 measurements is further complicated by matrix effects of the control materials [2].

References

  1. Lindemans J, Hoefkens P, van Kessel AL, Bonnay M, Kulpmann WR, van Suijlen JDE. Portable blood gas and electrolyte analyzer evaluated in a multiinstitutional study. Clin Chem 1999;45:111-117.
  2. Olafsdottir E, Westgard JO, Ehrmeyer SS, Fallon KD. Matrix effects on the performance and selection of QC procedures to monitor PO2 in blood gas measurements. Clin Chem 1996;42:392-6.