QC APPLICATION:
QC FOR
A POINT-OF-CARE GLUCOSE TEST:
A medical technologist who is a laboratory consultant recently
attended my course on the Selection, Evaluation, and Control
of Analytical Methods which is intended for senior CLS students
at UW-Madison. As a teacher, I enjoy having someone with several
years experience sit in on the course to stimulate discussion
and bring a real world perspective to the class. In discussing
her work, which includes consulting with labs on quality management,
she brought up a paper by Wandrup [1] that discussed the clinical
and analytical requirements and needs of glucose measurements
on whole blood. She wanted to see how the information in this
paper could be used with our QC planning process to select QC
procedures that would be appropriate for point-of-care (POC)
applications.
It's always interesting to take a paper from the literature
and show how this QC planning process should work using someone
else's numbers, so I used this information with an earlier paper
by Woo et. al [2] that presented method performance data on a
point-of-care analyzer. This made an especially good example
for the students because it brought together the recommendations
for quality from one paper and the estimates of method performance
from another to provide a more complete picture of how to evaluate
method performance and manage quality in a POC situation.
|
|
Note: this section of the WQC homepage is dedicated
to presenting example QC applications and illustrations of WQC
planning process. If you wish to submit your laboratory data
for free evaluation, please contact us by phone, fax or e-mail. |
|
We'll again follow the steps of the QC planning process.
1. Define the quality requirement.
In a section on analytical goal setting, Wandrup [1] provides
a statement of the clinical quality requirement in the following
way: "Clinicians sometimes indicate that measured differences
of 18-36 mg/dL (1-2 mmol/L) at a level of 90 mg/dL (5 mmol/L)
are therapeutically insignificant." This corresponds to what
we call a decision interval (Dint) of
(18/90)*100 = 20%
to (36/90)*100 = 40%.
2. Evaluate analytical and preanalytical
factors.
The imprecision of the glucose method at the decision level
of interest was found in Table 1 of the paper by Woo et. al [2]
where the CV was given as 2.44% at a control mean of 5.4 mmol/L.
This performance is based on data obtained when personnel in the
emergency department operated the device, so it should represent
performance under real operating conditions.
The inaccuracy of the method was determined from the regression
statistics presented by Woo et. al [2} in Table 3 (not shown here).
Based on 527 specimen comparisons between the POC analyzer and
a Beckman CX3 Analyzer, the regression line was given as y=-2.19
mg/dL + 1.051x. At a decision level of 90 mg/dL, the systematic
error can be calculated to be 2.4 mg/dL or 2.7%.
Within-subject biological variation is given by Wandrup as
an SD of 4.7 mg/dL, which would be
(4.7/90)*100 = 5.2% of the decision level of interest .
3. Enter parameters in computer program.
Because a clinical quality requirement is being used here,
it is necessary to account for the preanalytical within-subject
variation. Using the clinical model in the QC Validator program,
the parameters would be entered as shown here.
[For more information about entering parameters in the QC Validator program, download Tutorial D or the Program
Demo from the website.] |
|
4. Select a QC Procedure
As discussed by Woo et al [2], this portable clinical analyzer
would be subject to CLIA regulations, therefore 2 control materials
need to be analyzed to comply with the CLIA QC guidelines. With
the QC Validator 2.0's automatic QC selection, a QC procedure
can be selected by clicking the 2 Materials button. |
|
5. Review the selected QC procedure.
As shown by the OPSpecs chart generated by the program, a 12.5s
rule with N=2 will provide 90% detection of medically important
systematic errors, while having a 3% false rejection rate.
6. Document the performance of the selected
QC procedure.
In addition to the OPSpecs chart, it is useful to document
performance with critical-error graphs for both systematic error
and random error. The SE critical-error graph shows that a 13s/22s/R4s
multirule procedure will provide nearly as good error detection
as the 12.5s procedure and will have a little lower
false rejection rate. With the QC Validator program, you can
attach these charts to a summary page that documents all the
input paramters, calculated parameters, and your final QC selection. |
|
7. Adopt a total QC strategy
The Total QC strategy should include statistical QC, as well
as other QC components such as preventive maintenance, system
function checks (e.g., the electronic checks recommended by the
manufacturer), and periodic comparisons of performance with other
glucose measurement systems.
Because 90% AQA can be achieved here with an N of only 2,
you can rely on statistical QC and minimize other QC components.
It may still be worthwhile to improve method performance; e.g.
if method bias were reduced to 1%, then a 13s rule
could be used and a false rejection rate of 1% or less achieved
with a single rule QC procedure. |
|
In this table, the relative number of x's indicate the relative
emphasis on the different components in a Total QC strategy.
SQC stands for Statistical QC. QI stands for Quality Improvement.
Other QC includes preventive maintenance, instrument function
checks, preformance verification tests, and patient data QC algorithms.
|
|
8. Reassess for changes
It is interesting to assess how QC would be affected if the
bias of the method were eliminated. For example, if bias were
reduced from 2.7% to 0.0%, the OPSpecs chart here shows that
the method could be adequately controlled with 1 control measurement
per run using 3s control limits, i.e., 13s with N=1.
That's would be a simpler and less-costly control procedure. |
|
Conclusion: Consider a multistage QC strategy
If the bias can be eliminated (through better standardization
or calibration), then it would be cost-effective to implement
a two-stage QC strategy in which (a) 2 control materials could
be analyzed and interpreted with a 12.5s rule whenever
there are process changes, such as a new lot of reagents, a new
box of reagents brought into service after storage elsewhere,
or new less-experience operators, whereas (b) 1 control material
could be analyzed and interpreted with a 13s rule to
monitor performance during periods of stable operation.
It should be noted that the CLIA proficiency testing criterion
for acceptable performance is more demanding and would require
use of either 12.5s or 13s/22s/R4s/41s
with N=4 to assure that the allowable total error of 10% is not
exceeded. Thus, it would actually be desirable to check the test
more carefully when a new lot of reagents or major system change
occurs. The strategy to use a multistage QC procedure is still
applicable, but it may be better to use an N=4 QC procedure to
test for the effects of major system changes.
References
- Wandrup JH. Clinical and analytical requirements and needs
of glucose measurements on whole blood. Blood Gas News, 1996;5(number
3):3-8.
- Woo J, McCabe JB, Chauncey D, Schug T, Henry JB. The evaluation
of a portable clinical analyzer in the emergency department.
Am J Clin Pathol 1993;100:599-605.
We want your QC questions!
If you belong to a laboratory, hospital, or medical device manufacturers,
please e-mail us!
Copyright © 2000. All rights reserved.
Westgard QC, 7614 Gray Fox Trail, Madison WI 53717
Call 608-833-4718 or e-mail westgard@westgard.com
A Message from
JOW
QC Lessons | QC
Applications | Questions | Multirule
CLIA Requirements |
What's new? | Catalog
| Demo Download
Home | Glossary
| ARCHIVES | Links
| Feedback