An unofficial English translation of the RiliBÄK (Richtlinien der Bundesärztekammer). The term ‘RiliBÄK’ is an abbreviation meaning literally the Guidelines ("Rili") of the German Federal Medical Council (BÄK).
Posted October 2009
These guidelines include specifications for Acceptable % Root Mean Standard Deviation (RMSD) and Acceptable relative deviation for interlaboratory tests.
There are two types of target values listed for the interlaboratory test specifications. RMW means that a Reference Method Value was used to set the guideline. SW means that a target value specific for the test method (more like a peer group median) was used to set the guideline.
For more information on the meaning and use of these recommendations, read the accompanying guest essay by Jim Pierson-Perry and Oswald Sonntag.
Analytes in Plasma, Serum, and Whole Blood
| # |
Analyte |
Acceptable % RMSD |
Validity range of columns 3 and 5 |
Acceptable relative deviation in interlab tests |
Type of target value in interlab tests |
| lower limit |
upper limit |
units |
| 1 |
Activated partial thromboplastin time (aPTT) |
10.5% |
20 |
120 |
s |
18.0% |
SW |
| 2 |
Alanine aminotransferase (ALT) |
11.5% |
20 |
300 |
U/L |
21.0% |
RMW |
| 0.33 |
5.0 |
ukat/L |
| 3 |
Albumin |
12.5% |
20 |
70 |
g/L |
20.0% |
SW |
| 4 |
Alkaline phoshatase (AP) |
13.0% |
20 |
600 |
U/L |
21.0% |
SW |
| 0.33 |
10 |
ukat/L |
| 5 |
alpha-Fetoprotein(AFP) |
17.0% |
5 |
250 |
U/L |
24.0% |
SW |
| 6 |
Aspartate aminotransferase (AST) |
11.5% |
20 |
400 |
U/L |
21.0% |
RMW |
| 0.33 |
6.67 |
ukat/L |
| 7 |
Bilirubin (total) |
13.0% |
>2 |
30 |
mg/dL |
22.0% |
SW |
| >34 |
513 |
umol/L |
| 22.0% |
0.1 |
? 2 |
mg/dL |
| 1.7 |
? 34 |
umol/L |
| 8 |
Ca 19-9 |
14.0% |
5 |
300 |
KU/L |
27.0% |
SW |
| 9 |
Calcium (total) |
6.0% |
1 |
6 |
mmol/L |
10.0% |
RMW |
| 10 |
Calcium (ionized) |
7.5% |
>1 |
2.5 |
mmol/L |
15.0% |
SW |
| 14.5% |
0.2 |
? 1 |
mmol/L |
18.0% |
| 11 |
Carbamazepin |
12.0% |
2 |
20 |
mg/L |
20.0% |
SW |
| 12 |
Carcinoembryonic antigen (CEA) |
14.0% |
1 |
200 |
ug/L |
24.0% |
SW |
| 13 |
Chloride |
4.5% |
70 |
150 |
mmol/L |
8.0% |
RMW |
| 14 |
Cholesterin (total) |
7.0% |
50 |
350 |
mg/dL |
13.0% |
RMW |
| 1.3 |
9.1 |
mmol/L |
| 15 |
Cortisol |
16.0% |
>60 |
500 |
ug/L |
30.0% |
RMW |
| >166 |
1380 |
nmol/L |
| 18.5% |
20 |
60 |
ug/L |
| 55 |
? 166 |
nmol/L |
| 16 |
Creatine kinase (CK) |
11/0% |
50 |
1000 |
U/L |
20.0% |
RMW |
| 0.83 |
16.7+ |
ukat/L |
| 17 |
C-reactive protein (CRP) |
13.5% |
1 |
120 |
mg/L |
20.0% |
RMW |
| 18 |
Digitoxin |
15.5% |
5 |
80 |
ug/L |
30.0% |
RMW |
| 19 |
Digoxin |
14.0% |
>1 |
5 |
ug/L |
30.0% |
RMW |
| 17.5% |
0.5 |
? 1 |
ug/L |
| 20 |
Erythrocytes |
4.0% |
1.5 |
7 |
1012/L |
8.0% |
RMW |
| 21 |
Estradiol, 17-beta |
22.0% |
10 |
500 |
ng/L |
35.0% |
RMW |
| 37 |
1835 |
pmol/L |
| 22 |
Ethanol (clin. toxicological) |
9.0% |
>0.06 |
5 |
g/L |
12.0% |
SW |
| 15.0% |
0.2 |
? 0.6 |
g/L |
21.0% |
| 23 |
Ferritin |
13.5% |
10 |
600 |
ug/L |
25.0% |
SW |
| 24 |
Gamma-Glutamyl-Transferase (y-GT) |
11.5% |
20 |
300 |
U/L |
21.0% |
RMW |
| 0.33 |
5 |
ukat/L |
| 25 |
Glucose |
11.0% |
40 |
400 |
mg/dL |
15.0% |
RMW |
| 2.2 |
22 |
mmol/L |
| 26 |
Hematocrit |
5.0% |
10 |
60 |
% |
9.0% |
TV |
| 0.1 |
0.6 |
L/L |
| 27 |
Hemoglobin |
4.0% |
2 |
20 |
g/dL |
6.0% |
RMW |
| 1.2 |
12.4 |
mmol/L |
| 28 |
Hemoglobin A1c (HbA1c) Hemoglobin beta chain (Blood)-N-(1-deoxyfructos-1-yl) haemoglobin beta chain |
10.0% |
30 |
140 |
mmol/molHb |
18.0% |
RMV |
| 29 |
Uric acid |
7.0% |
2 |
13 |
mg/dL |
13.0% |
RMW |
| 119 |
773 |
umol/L |
| 30 |
Urea |
10.5% |
20 |
200 |
mg/dL |
20.0% |
RMV |
| 2.5 |
33 |
mmol/L |
| 31 |
Human chorionic gonadotropin (hCG) |
14.0% |
>100 |
1500 |
IU/L |
30.0% |
TV |
| 17.0% |
2 |
? 100 |
g/L |
| 32 |
Immunoglobulin A |
12.0% |
0.5 |
6 |
g/L |
20.0% |
SW |
| 33 |
Immunoglobulin G |
10.0% |
4 |
30 |
g/L |
180.0% |
SW |
| 34 |
Immunoglobulin MA |
13.0% |
0.4 |
5 |
g/L |
26.0% |
SW |
| 35 |
Potassium |
4.5% |
2 |
8 |
mmol/L |
8.0% |
RMW |
| 36 |
Creatinine |
11.5% |
0.5 |
10 |
mg/dL |
20.0% |
RMW |
| 44 |
884 |
umol/L |
| 37 |
Lactate |
11.0% |
9 |
90 |
mg/dL |
18.0% |
SW |
| 1 |
10 |
mmol/L |
| 38 |
Lactate dehydrogenase (LDH) |
9.0% |
100 |
700 |
U/L |
18.0% |
RMW |
| 1.67 |
11.7 |
ukat/L |
| 39 |
Leukocytes |
6.5% |
2 |
30 |
109/L |
18.0% |
RMV |
| 40 |
Lipase |
12.0% |
10 |
350 |
U/LM |
25.0% |
SW |
| 0.17 |
5.83 |
ukat/L |
| 41 |
Lithium |
6.0% |
0.3 |
3.5 |
mmol/L |
12.0% |
RMW |
| 42 |
Magnesium |
7.5% |
0.3 |
3.5 |
mmol/L |
15.0% |
RMW |
| 43 |
Sodium |
3.0% |
110 |
180 |
mmol/L |
5.0% |
RMW |
| 44 |
pCO2 |
6.5% |
15 |
110 |
mmHg |
12.0% |
SW |
| 45 |
pH |
0.4% |
6.75 |
7.80 |
|
0.80% |
RMW |
| 46 |
Pheonbarbital |
10% |
8 |
80 |
mg/L |
20.0% |
SW |
| 47 |
Phenytoin |
11.0% |
3 |
35 |
mg/L |
20.0% |
SW |
| 48 |
Phosphate (anorganic) |
9.0% |
1 |
10 |
mg/dL |
16.0% |
RMW |
| 0.3 |
3.2 |
mmol/L |
| 49 |
pO2 |
5.5% |
>125 |
350 |
mmHg |
12.0% |
SW |
| 7.0% |
>80 |
? 125 |
mmHg |
18.0% |
| 11.0% |
40 |
? 80 |
mmHg |
18.0% |
| 50 |
Progesterone |
17.0% |
>5.0 |
35 |
ug/L |
35.0% |
RMW |
| > 16 |
111 |
nmol/L |
| 22.0% |
0.2 |
? 5.0 |
ug/L |
| 0.6 |
? 16 |
nmol/L |
| 51 |
Prostate-specific antigen (PSA) |
15.5% |
0.2 |
50 |
ug/L |
25.0% |
SW |
| 52 |
Protein (Total) |
6.0% |
0.2 |
50 |
ug/L |
10.0% |
RMW |
| 53 |
Testosterone |
20.5% |
0.2 |
50 |
ug/L |
35.0% |
RMW |
| 0.7 |
69 |
nmol/L |
| 54 |
Theophylline |
13.0% |
3 |
40 |
mg/L |
24.0% |
RMW |
| 55 |
Thromboplastin time |
11.5% |
10 |
120 |
% |
23.0% |
SW |
| 56 |
Thrombocytes |
7.5% |
>300 |
700 |
109/L |
13.0% |
SW |
| 8.5% |
>150 |
? 300 |
109/L |
15.0% |
| 13.5% |
40 |
? 150 |
109/L |
18.0% |
| 57 |
Thyrotropic hormone (TSH) |
13.5% |
0.1 |
40 |
mU/L |
24.0% |
SW |
| 58 |
Thyroxine, total |
12.5% |
0.5 |
22 |
ug/L |
24.0% |
RMW |
| 6.4 |
283 |
nmol/L |
| 59 |
Thyroxine, free (FT4) |
14.0% |
>20 |
85 |
ng/L |
24.0% |
SW |
| >26 |
109 |
pmol/L |
| 15.5% |
2 |
? 20 |
ng/L |
| 2.6 |
? 26 |
pmol/L |
| 60 |
Transferrin |
9.5% |
0.5 |
6 |
g/L |
15.0% |
SW |
| 61 |
Triglycerides |
9.0% |
60 |
400 |
mg/dL |
16.0% |
RMW |
| 0.68 |
4.6 |
mmol/L |
| 62 |
Triiodine thyronine, total (T3) |
15.0% |
>1.2 |
10 |
ug/L |
24.0% |
SW |
| >1.8 |
15 |
nmol/L |
| 16.0% |
0.2 |
? 1.2 |
ug/L |
| 0.3 |
? 1.8 |
nmol/L |
| 63 |
Triidione thyronine, free (T3) |
14.5% |
1 |
25 |
ng/L |
24.0% |
SW |
| 1.5 |
39 |
pmol/L |
| 64 |
Troponin I |
20.0% |
0.1 |
35 |
ug/L |
33.0% |
SW |
| 65 |
Troponin T |
16.0% |
>1 |
8 |
ug/L |
33.0% |
SW |
| 21.0% |
0.08 |
? 1 |
ug/L |
| 66 |
Valproic acid |
11.5% |
20 |
150 |
mg/L |
20.0% |
SW |
| 67 |
Vancomycin |
13.0% |
>15 |
100 |
mg/L |
21.0% |
SW |
| 15.% |
4 |
? 15 |
mg/L |
Table B 1 b: Analytes in urine
| # |
Analyte |
Acceptable % RMSD |
Validity range of columns 3 and 5 |
Acceptable relative deviation in interlaboratory tests |
type of target value in interlaboratory tests |
| lower limit |
upper limit |
units |
| 1 |
Albumin |
15.0% |
1 |
500 |
mg/L |
26.0% |
SW |
| 2 |
Calcium |
8.5% |
0.5 |
6 |
mmol/L |
17.0% |
SW |
| 3 |
Glucose |
11.0% |
100 |
4000 |
mg/L |
22.0% |
RMW |
| 0.6 |
22 |
mmol/L |
| 4 |
Uric acid |
13.5% |
5 |
300 |
mg/L |
23.0% |
RMW |
| 30 |
1784 |
umol/L |
| 5 |
Urea |
13.5% |
0.1 |
20 |
g/L |
21.0% |
RMW |
| 1.7 |
333 |
mmol/L |
| 6 |
Potassium |
8.5% |
2 |
140 |
mmol/L |
15.0% |
RMW |
| 7 |
Creatinin |
12.0% |
0.01 |
3 |
g/L |
21.0% |
RMW |
| 0.1 |
27 |
mmol/L |
| 8 |
Sodium |
6.5% |
50 |
200 |
mmol/L |
12.0% |
RMW |
| 9 |
Phosphate (anorganic) |
11.5% |
30 |
900 |
mg/L |
20.0% |
SW |
| 1 |
29 |
mmol/L |
| 10 |
Protein (total) |
11.5% |
5 |
10000 |
mg/L |
24.0% |
SW |
Table B 1 c: Analytes in cerebrospinal fluid
| # |
Analyte |
Acceptable % RMSD |
Validity range of columns 3 and 5 |
Acceptable relative deviation in interlaboratory tests |
Type of target value in interlaboratory tests |
| lower limit |
upper limit |
units |
| 1 |
Albumin |
13.5% |
20 |
1000 |
mg/L |
23.0% |
SW |
| 2 |
Glucose |
9.5% |
20 |
300 |
mg/dL |
18.0% |
RMW |
| 1.1 |
17 |
mmol/L |
| 3 |
Immunoglobulin A |
15.5% |
2 |
40 |
mg/L |
27.0% |
SW |
| 4 |
Immunoglobulin G |
12.0% |
15 |
500 |
mg/L |
20.0% |
SW |
| 5 |
Immunoglobulin M |
15.5% |
1 |
30 |
mg/L |
33.0% |
SW |
| 6 |
Lactate |
11.5% |
10 |
99 |
mg/dL |
20.0% |
SW |
| 1.1 |
11 |
mmol/L |
| 7 |
Protein (total) |
13.5% |
10 |
2000 |
mg/L |
23.0% |
SW |
|