Using Allowable Total Error in Assay Migration - Westgard QC

Tools, Technologies and Training for Healthcare Laboratories

Using Allowable Total Error in Assay Migration

Total Allowable Error can be used to benchmark test performance. It can be used to help optimize QC Design and determine Sigma-metrics. But can it be helpful when moving tests from one platform to another? One of our most popular guest authors, Hans van Schaik, discusses how he used TEa in this way at Ziekenhuis St. Antonius

Using Total Error allowable in the migration of immunochemistry.

MAY 2013

Hans van Schaik, Core Analyst quality assurance and process control
Department of clinical chemistry
Sint Antoniusziekenhuis
Nieuwegein, The Netherlands

 As Siemens announced the end of the technical support for our two (former DPC) Immulite 2500 analysers was due for july 2012, it was decided to move the immunochemistry tests performed on these machines to the Roche Cobas 6000-CEE analyser.

Only a limited number of tests that were not available from Roche (yet) and the allergy testing would be installed on the replacing Immulite 2000 XPI.

Reagents for the Immulite 2500 would be available from Siemens until june 2012, as long as stock lasted.

For this reason it was decided to start implementing the tumor markers (AFP, CEA, total PSA, free PSA, CA19.9 and CA125) first, followed by ß2-microglobulin, Ferritin, FSH, Progesterone, Testosterone, PTH, T3 and SHBG.

CA19.9 was not performed on the Immulite 2500, but sent to an external laboratory, it was decided to implement this test on the Cobas 6000 too.

CA15.3 has been installed on the Cobas 6000 before and is therefore not in the comparison.

ß2-microglobulin is performed on the C-module of the Cobas 6000 and will not be included in this essay.

The migration had to be planned carefully, because immunochemistry is far less standardised and harmonised, causing differences in results produced on different systems.

Especially for the tumor markers these method based differences can cause problems in the follow-up of patient results.

 Before starting the evaluation the EQA results were studied, giving us a first impression about the differences we could expect.

The calculated differences are shown in table 1.

 HVS-Table1

  For the method comparison of the tests two different procedures were used:

-       To avoid the influence of different epitopes in the tumormarker assay we made a low and high pooled serum, each composed of at least 20 different samples. Both pools were measured in 5-fold on the Immulite and Cobas 6000.

-       The other tests were evaluated by measuring 20 individual patient samples on both analysers.

-       Slope and offset were calculated.

CA19.9 was compared by checking the result received from the external lab with the result produced on the Cobas 6000.

The results are shown in table 2:

 HVS-Table2

 The results in table 2 confirm the differences seen in the EQA scheme.

Because of the variability in results for the tumor markers it was decided to measure patient samples on both analysers for a year, only for those patients who already had historical results produced with the Siemens method. Samples from patients without historical results were analysed with the Roche method only. Once a patient had a Cobas result in the history the follow-up was continued with the Roche method.

This procedure was programmed in our laboratory information system, eventually triggering the drawing of an extra tube of blood for the measurement on the Immulite.

Because the period of measuring tumor markers on both Immulite and Cobas 6000 would exceed the availability of reagents for the Immulite 2500, the tumor markers were temporary installed on the Immulite 2000 XPI.

It is expected that in the end of 2012 the double measurement will come to an end and all samples can be measured on the Cobas 6000 only.

 The allowable percentage of difference in the results with both methods was determined based on TEa, as shown in table 3:

HVS-Table3 

 For the samples measured with both analysers the ratio between Immulite and Cobas 6000 result was calculated.

When de ratio was < 0,80 or > 1,20 the Cobas 6000 result was reported with a remark showing the Immulite result: [Testname] previous method = ……. [unit] (method changed 01-03-2012).

The historical results were checked and in case of unexpected differences measurements were repeated.

 Setting the QC-rules using TEa

 For setting the QC-rules for most tests we used the long term CV (>200 results) from our internal QC and assumed the bias 0. Because of a QC-lot change since the migration we only had 42 – 49 results available for some tests.

For the TEa the desirable specifications from the “Ricos table” were used.

These numbers were entered in Westgard QC’s EZ Rules3 program, giving us an advice on the set of

qc-rules to apply and the number and frequency of control measurements.

 Table 4 shows the Sigma-scores for each level, the number of results used to calculate the sigmascore (n) and the QC-Rules and frequency advised by EZRules3.

 HVS-Table4

 If for level 1 and 2 different QC-rules were advised the most stringent set of rules was applied, except for ferritin, where the high control has a value in a clinical not relevant range.

 This project has been performed with the cooperation of Netty van Trooyen, supervising analyst immunochemistry, Ineke Dijkstra, clinical chemist in residence and Floris de With, supervising analyst clinical chemistry.

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input