Tools, Technologies and Training for Healthcare Laboratories

Laboratory INVITRO Sigma Verification of Performance

In 2016, INVITRO Laboratories became the first (and largest) private laboratory in Russia to achieve the Verification of Performance. (Sigma VP). More than 40 analytes of two of their chemistry instruments were verified. In 2017, they re-verified this performance. In 2018, they expanded their verification to 4 other labs across Russia, encompassing 129 assays.

INVITRO Laboratories - Sigma Verification of Performance

Verified since September 27th, 2016
Duration of Verification: November 1st 2019 through December 31st 2020

In 2016, Westgard QC was pleased to verify the performance of more than 45 assays at INVITRO, the largest private medical company in Russia.

 INVITRO announcement

INVITRO has 9 modern labs and owns the biggest network in Eastern Europe with over 900 medical offices in Russia, Ukraine, Kazakhstan, and Belarus. It was the first laboratory in Russia to obtain ISO 15189:2012 certification and ISO 9001:2011 (ISO 9001:2008). It participates in international external quality assurance programs that include FSVOK (Russia), EQAS - USA, and RIQAS - UK.

This pilot verification assessed the performance of two of their main core laboratory biochemistry instruments, the c8000 mid-volume chemistry analyzer and  the c16000 high volume chemistry analyzer

In 2018, INVITRO expanded their pilot programs to 4 other branches in their network, further expanding to 5 branches in 2019.

List of analytes verified

Chemistry Assays

Moscow

C1601772 Verified

C1601770 Verified

Albumin

Six Sigma

Six Sigma

Alkaline Phosphatase

Six Sigma

Six Sigma

ALT

Six Sigma

Six Sigma

Amylase

Six Sigma

Six Sigma

AST

Six Sigma

Five Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Bilirubin, Total

Six Sigma

Six Sigma

Calcium

Six Sigma

Six Sigma

Chloride

Five Sigma

Four Sigma

Cholesterol

Six Sigma

Five Sigma

Creatinine Kinase

Six Sigma

Six Sigma

Creatinine

Six Sigma

Six Sigma

GGT

Four Sigma

Six Sigma

Glucose

Four Sigma

Five Sigma

HDL

Five Sigma

Six Sigma

Iron

Five Sigma

Four Sigma

LDH

Four Sigma

Four Sigma

Magnesium

Five Sigma

Six Sigma

Phosphorous

Six Sigma

Six Sigma

Potassium

Six Sigma

Six Sigma

Protein, Total

Six Sigma

Six Sigma

Sodium

Four Sigma

-

Triglycerides

Six Sigma

Six Sigma

Uric Acid

Six Sigma

Six Sigma

 

Urine Assays

Moscow

C1601770 Verified

Amylase, U

Six Sigma

Calcium, U

Six Sigma

Glucose, U

Six Sigma

Magnesium, U

Six Sigma

Phosphorous, U

Six Sigma

Potassium, U

Six Sigma

Protein, U

Six Sigma

Sodium, U

Six Sigma

Urea N, U

Six Sigma

Uric, U

Six Sigma

 

Immunoassays

Moscow

AI23004 Verified

AI23005 Verified

Cortisol

Four Sigma

-

Estradiol

Five Sigma

-

FSH

Five Sigma

Five Sigma

LH

Six Sigma

-

Progesterone

Six Sigma

-

Prolactin

Six Sigma

-

Testosterone

Six Sigma

Six Sigma

 

 

Chemistry Assays

Siberia

C804275 Verified

Albumin

Six Sigma

Alkaline Phosphatase

Six Sigma

ALT

Six Sigma

Amylase

Six Sigma

AST

Five Sigma

Bilirubin, Direct

Six Sigma

Bilirubin, Total

Six Sigma

Calcium

Six Sigma

Chloride

Five Sigma

Cholesterol

Six Sigma

Creatinine Kinase

Six Sigma

Creatinine

Six Sigma

GGT

Six Sigma

Glucose

Six Sigma

HDL

Six Sigma

Iron

Four Sigma

LDL

Six Sigma

Lipase

Six Sigma

Magnesium

Five Sigma

Phosphorous

Six Sigma

Potassium

Six Sigma

Protein, Total

Six Sigma

Transferin

Six Sigma

Triglycerides

Six Sigma

UIBC

Five Sigma

Urea Nitrogen

Five Sigma

Uric Acid

Six Sigma

 

Immunoassays

Siberia

ISR61139 Verified

ISR61143 Verified

AI22405 Verified

Anti-TG

-

-

Five Sigma

Anti-TPO

-

-

Six Sigma

AFP

-

Six Sigma

-

Estradiol

Six Sigma

-

-

Folate

Four Sigma

-

-

LH

Four Sigma

-

-

Progesterone

Six Sigma

-

-

Prolactin

-

Six Sigma

-

iPTH

-

-

Five Sigma

T3, Total

-

-

Four Sigma

T4, Free

-

-

Five Sigma

T4, Total

-

-

Five Sigma

Testosterone

-

Five Sigma

-

 

Chemistry Assays

Ural

804573 Verified

C804891 Verified

Albumin

Six Sigma

Six Sigma

Alkaline Phosphatase

Six Sigma

Six Sigma

ALT

Six Sigma

Six Sigma

Amylase

Six Sigma

Six Sigma

AST

Four Sigma

Four Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Bilirubin, Total

Six Sigma

Five Sigma

C3

Six Sigma

-

C4

Six Sigma

-

C-Reactive Protein

Six Sigma

-

Calcium

Five Sigma

Five Sigma

Chloride

Four Sigma

Five Sigma

Cholesterol

Four Sigma

Six Sigma

CO2

-

Four Sigma

Creatinine Kinase

Six Sigma

-

Creatinine

Six Sigma

Six Sigma

GGT

Six Sigma

Six Sigma

Glucose

Four Sigma

Four Sigma

HDL

Six Sigma

Six Sigma

IgA

Six Sigma

-

IgG

Six Sigma

-

IgM

Six Sigma

-

Iron

Six Sigma

Five Sigma

Lactate

Four Sigma

-

LDL

Six Sigma

-

LDH

-

Five Sigma

Lipase

Five Sigma

-

Magnesium

-

Five Sigma

Phosphorous

Six Sigma

Six Sigma

Potassium

Six Sigma

Six Sigma

Protein, Total

Six Sigma

Six Sigma

Sodium

-

Four Sigma

Triglycerides

Six Sigma

Six Sigma

UIBC

Four Sigma

-

Uric Acid

Four Sigma

Five Sigma

 

 

Urine Assays

Ural

C804891 Verified

Amylase, U

Six Sigma

Calcium, U

Six Sigma

Creatinine, U

Six Sigma

Glucose, U

Six Sigma

Magnesium, U

Six Sigma

Phosphorous, U

Six Sigma

Protein, U

Six Sigma

Urea N, U

Six Sigma

Uric, U

Six Sigma

 

Immunoassays

Ural

AI05057 Verified

AI21777 Verified

AFP

-

Four Sigma

FSH

Five Sigma

-

LH

Five Sigma

-

Progesterone

Five Sigma

-

T4, Free

Five Sigma

-

T4, Total

Five Sigma

-

Prolactin

-

Six Sigma

PSA

-

Five Sigma

Testosterone

Six Sigma

-

TSH

Six Sigma

-

 

Tumor Markers assays

Ural

AI21777 Verified

CA 125 II

Four Sigma

CA 15-3

Four Sigma

CA 19-9 XR

Four Sigma

 

Chemistry Assays

Kazakhstan

C804691 Verified

Albumin

Six Sigma

Alkaline Phosphatase

Five Sigma

ALT

Six Sigma

Amylase

Six Sigma

AST

Five Sigma

Bilirubin, Direct

Six Sigma

Bilirubin, Total

Six Sigma

Calcium

Four Sigma

Chloride

Four Sigma

Creatinine

Six Sigma

GGT

Six Sigma

Glucose

Six Sigma

HDL

Five Sigma

Iron

Six Sigma

LDL

Six Sigma

Magnesium

Five Sigma

Potassium

Six Sigma

Protein, Total

Six Sigma

Triglycerides

Six Sigma

Urea Nitrogen

Six Sigma

Uric Acid

Six Sigma

 

Immunoassays

Kazakhstan

I63209 Verified

I63210 Verified

Anti-TG

-

Six Sigma

Anti-TPO

-

Six Sigma

Estradiol

Five Sigma

-

Ferritin

Six Sigma

-

Folate

Six Sigma

-

FSH

Five Sigma

-

Insulin

Six Sigma

-

Progesterone

Four Sigma

-

Prolactin

Six Sigma

-

T3, Free

-

Four Sigma

T3, Total

-

Four Sigma

T4, Free

-

Six Sigma

T4, Total

-

Four Sigma

Testosterone

Six Sigma

-

TSH

-

Six Sigma

 

Tumor Markers assays

Kazakhstan

I63209 Verified

CA 125 II

Six Sigma

CA 15-3

Four Sigma

CA 19-9 XR

Five Sigma

 

Chemistry Assays

Belarus

C804526 Verified

C804627 Verified

Albumin

-

Six Sigma

Alkaline Phosphatase

Six Sigma

Six Sigma

ALT

Six Sigma

Six Sigma

Amylase

-

Six Sigma

AST

Six Sigma

Five Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Bilirubin, Total

Six Sigma

Six Sigma

C-Reactive Protein

Six Sigma

Six Sigma

Calcium

-

Four Sigma

Chloride

Four Sigma

-

Cholesterol

Six Sigma

Six Sigma

Creatinine Kinase

-

Six Sigma

Creatinine

Six Sigma

Six Sigma

GGT

Six Sigma

Six Sigma

Glucose

Four Sigma

Four Sigma

HDL

Four Sigma

Six Sigma

Iron

-

Five Sigma

LDL

Six Sigma

-

LDH

-

Four Sigma

Lipase

-

Six Sigma

Magnesium

-

Six Sigma

Phosphorous

-

Six Sigma

Potassium

Six Sigma

-

Protein, Total

Five Sigma

Four Sigma

Transferin

-

Six Sigma

Triglycerides

Six Sigma

Six Sigma

Uric Acid

-

Six Sigma

 

Urine Assays

Belarus

C804526 Verified

C804627 Verified

Creatinine, U

-

Six Sigma

Glucose, U

-

Six Sigma

Protein, U

Six Sigma

-

Uric, U

-

Six Sigma

 

Immunoassays

Belarus

I62718 Verified

I62753 Verified

Cortisol

Five Sigma

-

Estradiol

Six Sigma

-

Ferritin

-

Four Sigma

Folate

-

Six Sigma

Insulin

-

Six Sigma

LH

Four Sigma

-

Progesterone

Four Sigma

-

Prolactin

Four Sigma

-

T3, Free

-

Four Sigma

T3, Total

-

Four Sigma

T4, Free

-

Six Sigma

T4, Total

-

Six Sigma

Testosterone

Six Sigma

-

Vitamin B12

-

Four Sigma

 

 

Chemistry Assays

St. Petersburg

C804721 Verified

Albumin

Six Sigma

Alkaline Phosphatase

Six Sigma

ALT

Six Sigma

Amylase

Six Sigma

AST

Six Sigma

Bilirubin, Direct

Six Sigma

Bilirubin, Total

Six Sigma

C-Reactive Protein

Six Sigma

Calcium

Six Sigma

Chloride

Five Sigma

Cholesterol

Six Sigma

Creatinine Kinase

Six Sigma

Creatinine

Six Sigma

GGT

Six Sigma

Glucose

Five Sigma

HDL

Six Sigma

IgA

Six Sigma

IgG

Six Sigma

IgM

Six Sigma

Iron

Four Sigma

LDL

Four Sigma

LDH

Four Sigma

Lipase

Five Sigma

Magnesium

Five Sigma

Phosphorous

Six Sigma

Potassium

Six Sigma

Protein, Total

Six Sigma

Sodium

Four Sigma

Transferin

Six Sigma

Triglycerides

Five Sigma

UIBC

Six Sigma

Urea Nitrogen

Four Sigma

Uric Acid

Six Sigma

 

Urine Assays

St. Petersburg

C804721 Verified

Creatinine, U

Six Sigma

Glucose, U

Six Sigma

Protein, U

Six Sigma

Uric, U

Six Sigma

 

Immunoassays

St. Petersburg

AI04567 Verified

ISR63357 Verified

ISR56113 Verified

Anti-TG

Four Sigma

-

-

Anti-TPO

Six Sigma

-

-

AFP

-

-

Five Sigma

Cortisol

-

-

Five Sigma

Estradiol

-

Six Sigma

-

Ferritin

-

-

Four Sigma

Folate

-

-

Four Sigma

FSH

-

Five Sigma

-

Insulin

-

-

Six Sigma

LH

-

Five Sigma

-

Progesterone

-

Six Sigma

-

Prolactin

-

Six Sigma

-

iPTH

Five Sigma

-

-

T3, Free

Five Sigma

-

-

T3, Total

Four Sigma

-

-

T4, Free

Six Sigma

-

-

T4, Total

Five Sigma

-

-

Testosterone

-

Six Sigma

-

TSH

Six Sigma

-

-

Vitamin B12

-

-

Five Sigma

 

Tumor Markers assays

St. Petersburg

AI04567 Verified

CA 125 II

Four Sigma

CA 15-3

Four Sigma

CA 19-9 XR

Four Sigma

 

Chemistry Assays

Samara

C802571 Verified

Alkaline Phosphatase

Five Sigma

ALT

Six Sigma

Amylase

Five Sigma

AST

Four Sigma

Bilirubin, Direct

Six Sigma

Bilirubin, Total

Five Sigma

C-Reactive Protein

Six Sigma

Cholesterol

Four Sigma

Creatinine

Six Sigma

HDL

Six Sigma

Iron

Five Sigma

LDL

Four Sigma

Phosphorous

Six Sigma

Transferin

Six Sigma

Triglycerides

Four Sigma

Immunoassays

Samara

I60295 Verified

AI04150 Verified

Anti-TPO

-

Six Sigma

Ferritin

-

Four Sigma

FSH

Four Sigma

-

Insulin

-

Six Sigma

Progesterone

Four Sigma

-

Prolactin

Six Sigma

-

Testosterone

Six Sigma

-

TSH

-

Six Sigma

Vitamin B12

-

-

Visual Display of Analytical Performance of INVITRO

2023 June INVITRO NMEDX 1

2023 June INVITRO NMEDX 2

2023 June INVITRO NMEDX 3 

 2023 June INVITRO NMEDX 4

2023 June INVITRO NMEDX 5

2023 June INVITRO NMEDX 6

2023 June INVITRO NMEDX 7

2023 June INVITRO NMEDX 8

2023 June INVITRO NMEDX 9

2023 June INVITRO NMEDX 10

2023 June INVITRO NMEDX 11

2023 June INVITRO NMEDX 12

2023 June INVITRO NMEDX 14

2023 June INVITRO NMEDX 152023 June INVITRO NMEDX 16

2023 June INVITRO NMEDX 17

2023 June INVITRO NMEDX 18

2023 June INVITRO NMEDX 19

2023 June INVITRO NMEDX 20

2023 June INVITRO NMEDX 21

2023 June INVITRO NMEDX 22

2023 June INVITRO NMEDX 23

2023 June INVITRO NMEDX 24

2023 June INVITRO NMEDX 25

2023 June INVITRO NMEDX 26

2023 June INVITRO NMEDX 27

2023 June INVITRO NMEDX 28

2023 June INVITRO NMEDX 29

2023 June INVITRO NMEDX 30

2023 June INVITRO NMEDX 31

2023 June INVITRO NMEDX 32

2023 June INVITRO NMEDX 33

2023 June INVITRO NMEDX 34

2023 June INVITRO NMEDX 35

2023 June INVITRO NMEDX 36

2023 June INVITRO NMEDX 37

 

 

 

 

 

Congratulations to INVITRO on their accomplishments.

INVITRO Verification

Sten Westgard, MS, pictured here in 2016 with INVITRO founder and Chairman of the Board, Alexander Ostrovskiy.

 INVITRO Elena Kondrashova

Pictured here on the right is Elena Kondrasheva, Laboratory Director.