Questions and Answers - Westgard QC - Blog - Page 2 - Results from #20

Tools, Technologies and Training for Healthcare Laboratories

Q & A: The Rong Regulation, Turkey Mis-Totaling Total Error

Posted by Sten Westgard, MS

Often we get questions about the problems that laboratories encounter when they wrestle with their local regulations.

Here we have a case where there are Turkish laboratory regulations that are irrational. In the words of the submitter:

"I have a question about the estimation of total CV. We have a requlation about total allowable error in Turkey....It is Turkish, but I strongly believe that you can follow it.

"The marked page can be seen below: There is an example for glucose. Two levels of controls are measured and the CV%’s calculated. The total CV are calculated from the CV’s of two different levels as seen in the equation. I can not accept it, because the levels are different even if the CV’s are close."

TurkeyTotalErrorCombination

What's gone Rong here? More after the jump...

-----
Continue reading
  821 Hits

Q&A: 10 Questions on QC from the first webinar

Posted by Sten Westgard, MS

Continue reading
  2045 Hits