I came across a troubling admission in an exchange of letters about a study of the quality of POC glucose meters:

"....In our study, the meters had a CV of 2.5-5%. A glucose level of 7.0 mmol/L could thus read anywhere between 6.3 and 7.7 mmol/L with 95% confidence. With this degree of variance, single readings using these meters should not be used for either the diagnosis of diabetes, or for making critical clinical decisions, e.g. initiating insulin. Glucose meters are best used by analysing the results of multiple readings over time, which reduces the error involved, rather than relying on single readings...."

Matthew Cohen, Jonathan Shaw, Response to the letter by Cheung et al., Diabetes Research and Clinical Practice 72 (2006) 218

Maybe I'm missing something, but wasn't the point of POC devices to get quick, accurate readings you can act upon? If the only way to get useful results out of POC meters is to make multiple readings over time, why not just order the lab test? You already have to wait - so why not get a good, accurate single result from the better method?

It's my naivete, I'm sure.

-----

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Video: "Westgard Rules"
What Doctors Don't Know about What They Don't Know
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Thursday, 22 October 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input