Tools, Technologies and Training for Healthcare Laboratories
  • Home

Sigma Verification of Saigon Hoan My Performance

The laboratory of the Hoan My Sai Gon hospital in Ho Chi Minh City, Vietnam has joined the Sigma VP program.

Sigma Verification of the Laboratory of the Hoan My Sai Gon Hospital

Verification: December 12th, 2017
Duration of Verification: December 12th, 2017 through January 1st, 2019

"Hoan My Sai Gon (HMSG) hospital is the largest site of Hoan My Medical system which is recognized as the biggest private hospital chain in Vietnam. Hoanmy system includes 11 hospitals and 4 clinics and continue to enlarge the network with many M&A and expansion projects. The laboratory of HMSG hospital plays a very important role as it is the biggest LAB with leading quality positioning and commitment within the network. HMSG LAB was accredited for ISO 15189 certification in May, 2017. Everyday, this LAB serves around 1000 patients with more than 10,000 test of clinical chemistry, immunology, hematology, microbiology and molecular biology. HMSG LAB always keep improving the quality in order to serve their customers better."

Analyte

c803812  verified performance

c803866  verified performance

Albumin

 

Six Sigma

Amylase

Six Sigma

Six Sigma

ALT

Six Sigma

Six Sigma

AST

Six Sigma

Six Sigma

Bilirubin, Direct

Six Sigma

Six Sigma

Bilirubin, Total

Six Sigma

Six Sigma

C-reactive Protein

Six Sigma

Six Sigma

Calcium

Five Sigma

Five Sigma

Chloride

Six Sigma

Four Sigma

Cholesterol, total

Five Sigma

Six Sigma

Cholesterol, HDL

Six Sigma

Six Sigma

Cholesterol, LDL

Four Sigma

Four Sigma

Creatinine

Six Sigma

Six Sigma

GGT

Six Sigma

Six Sigma

Glucose

Six Sigma

Five Sigma

LDH

 

Six Sigma

Magnesium

   

Potassium

Six Sigma

Six Sigma

Total Protein

 

Four Sigma

Triglycerides

Six Sigma

Six Sigma

Urea Nitrogen

Five Sigma

Five Sigma

Uric Acid

Six Sigma

Six Sigma

 The assays and instruments which were verified

 Six Sigma: World Class Quality. Possible to use 3s to 3.5s limits and a minimum of controls

Five Sigma: Excellent Quality. Possible to use 3s limits or a 1:3s/2:2s/R:4s Westgard Rule

Four Sigma: Good Quality. Possible to use 1:3s/2:2s/R:4s/4:1s Westgard Rule

 Visual Summary of Performance of Hoan My Sai Gon Hospital Lab

 2017 SigmaVP Hoanmy c8000 1 NMEDX

 2017 SigmaVP Hoanmy c8000 2 NMEDX

 

Congratulations to the entire laboratory staff of the Hoan My Sai Gon hospital.

Joomla SEF URLs by Artio