The Clinical Laboratory of the Women's Hospital of the School of Medicine of Zheijiang University, in Hangzhou, China has joined the Sigma Verification of Performance program
Assay |
C1600418 |
C1600757 |
C1600418 |
Alkaline Phosphatase |
Six Sigma |
Six Sigma |
Six Sigma |
AST |
Six Sigma |
Six Sigma |
Six Sigma |
Direct Bilirubin |
Six Sigma |
Six Sigma |
Six Sigma |
Total Bilirubin |
Four Sigma |
Four Sigma |
Six Sigma |
C-Reactive Protein |
Six Sigma |
Six Sigma |
Six Sigma |
Chloride |
Five Sigma |
Six Sigma |
Six Sigma |
GGT |
Six Sigma |
Six Sigma |
Six Sigma |
Creatinine Kinase |
Six Sigma |
Six Sigma |
Six Sigma |
GGT |
Six Sigma |
Six Sigma |
Six Sigma |
Potassium |
Six Sigma |
Six Sigma |
Six Sigma |
Sodium |
Five Sigma |
||
Urea Nitrogen |
Four Sigma |
Four Sigma |
Six Sigma: World Class Quality. Possible to use 3s to 3.5s limits and a minimum of controls
Five Sigma: Excellent Quality. Possible to use 3s limits or a 1:3s/2:2s/R:4s Westgard Rule
Four Sigma: Good Quality. Possible to use 1:3s/2:2s/R:4s/4:1s Westgard Rule