After premiering this event at ADLM, we are opening it up to everyone through zoom. Please join us September 25th, at 11 am central standard time. At last, a frank look at why we need to do less QC, and how we can make those reductions safely.  If you enjoy being out of control every day, if you get a thrill every time you have to repeat a con...
Q and A: Establishing a new mean and SD within the manufacturer's range
Sten Westgard
Uncategorized
"I have a question about establishing a Levey-Jennings control chart. laboratories should collect a minimum of 20 measurements over at least 10 days to calculate the mean and SD for the chart. My question is: Is it required that these initial 20 data points must fall within the manufacturer's declared range (or assigned values) for the control...
Should Labs impose Desperation Premiums?
Sten Westgard
Uncategorized
Every so often it's useful to transpose a few practices from outside the laboratory into the laboratory, just to see how we would evaluate them. The new Tech powers are gleefully implementing a system called "surveillance pricing", which they sell as a dynamic way to sell things to customers, except it turns out that they really only want the prici...
A solution to IVDR over-regulation?
Sten Westgard
Current Affairs
 The fallout of the IDVR continues. An interesting quote from a recent paper: "Although the EC aimed at bringing safe and effective medical tests to the EU-market, ~9,000 tests out of ~40,000 tests got lost in transition." Nearly a quarter of tests suddenly became unacceptable? That places IVDR at just above 2.2 Sigma. The proposed soluti...
Indonesian hematology analyzer that can't hit anything
Sten Westgard
Uncategorized
A paper from Surabaya, Indonesia assessed the performance of an anonymous hematology analyzer. The study used the old CLIA 1992 goals. We updated it with an assessment using CLIA 2025 goals. And then it became clear why this instrument was not named. None of this performance would meet CLIA's standards. Is performance so bad in this analyzer, ...
Question and Answer: How to use the 7:T rule?
Sten Westgard
Uncategorized
This question recently came in: "We have two specific questions regarding the practical application of the 7T rule: "When applying this rule, is it mandatory to use the 1:2s rule as a starting trigger, or can the laboratory choose to use 7T independently without the 1:2s initiation? (Personally, I believe 1:2s initiation is necessary….) "When a 7T ...
4:1s, when is it violated under different types of Westgard Rules?
Sten Westgard
Uncategorized
The Westgard Rules, in the classic 1981 multirule paper, require a 1:2s violation before other rejection rules are violated.  If you had4 values past 1s, but nothing past 2s, it wouldn't be checked. Under more "modern" Westgard Rules, with the 1:2s warning rule abolished, you don't need to wait until the 2s limit line is violated before interp...
Repeating the control: just more info, or just rolling the dice?
Sten Westgard
Uncategorized
A recent question popped up in one of our recent webinars: "It seems to me that repeating a QC may or may not have value, yes if it provides further troubleshooting data, and not if it is used to override a proper error flag. But to group the two together, without making that distinction seems inappropriate. Perhaps expand the survey ques...
Q and A: Is the 10:x rule still compulsory?
Sten Westgard
Uncategorized
A recent question came to me via social media: "Hello Sir, Good morning, How are you? I wanted to know is still 10x rule is coumpulsary?[stet.]  Even [when] we are using lab mean and lab SD. Please share the details.Thanks and Regards, [anonymous]"10:x is not compulsory by any regulation that I am aware of. Many regulations require you to...
"Real" analytical Sigma metrics, a rarity
Sten Westgard
Uncategorized
[Full disclosure, Westgard QC has a partnership with Beckman Coulter. However, this paper was not prepared by Westgard, but was written completely independently. We are simply calling attention to the study's unique approach and findings.] A new study from Tamil Nadu, India, has gone into unprecedented detail of the analytical performance. Typicall...