Q & A: We don't run QC with a time machine

A question came in about running QC and looking at the initial 20 data points used for validation. Full Question: "I want to ask: when running in parallel, the first 20 data points don't have Westgard rules applied, and the data is within the manufacturer's 2SD limit. "However, once the laboratory mean/SD is calculated and the laboratory&...

Continue reading
  62 Hits

Q: What if we can't Westgard Sigma Rules?

Another Question from the "Cut Your QC" webinar: "If management is unwilling to adopt the Sigma-metric method, is there a better Westgard Rule than 1:2s to apply across the board to all analytes?" If Westgard Sigma Rules aren't a possibility at your laboratory, there are still better ways to run QC using the 1:2s rule. Remember, with 2 controls, th...

Continue reading
  169 Hits

Q: For every method, instrument, reagent lot, a new QC rule?

Q: "You mentioned that there can be variation in performance even with different reagent lots. Are you recommending that we change QC rules based on the performance of each lot on each instrument?" This is another question from our Cut Your QC webinar, now available as an on-demand recording. https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6e...

Continue reading
  310 Hits

10 of 27 analytes on a Cobas Pure cannot hit desirable MAU

I came across a rare but intriguing study of measurement uncertainty recently:Standardized Estimation of Measurement Uncertainty in Clinical Chemistry Using ISO 20914. Mohammad I. Ahmad, Fayha S Ahmed, Deepa Rao, Mohamedi Begum, Sheifa Joan. Clin Lab 2026;72:1-6. https://www.clin-lab-publications.com/article/5788The study looked at 27 general bioch...

Continue reading
  212 Hits

Where do all the errors happen?

I read a recent paper on errors across the Total Testing Process (TTP) recently. It's a unique paper, one that delved into the types of non-conformances.   Evaluation of Non Conformances in the Total Testing Process of a NABL-accredited Clinical Biochemistry Laboraotry: An Observational Study. Dharmikkumar Savjibhai Patel, Harshkumar Darj...

Continue reading
  190 Hits

Q: Can POC Cut their QC?

"Can this approach [Cut Your QC] be used on QC for Point of Care devices? Are rules built into these devices that can't be modified like standard central lab instruments?"This is another question from our Cut Your QC webinar, now available as an on-demand recording. https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQThe questioner sa...

Continue reading
  237 Hits

Q: 1 level is out 3 sd, but the other 2 levels are "in", what do we do?

"We have a 1:3s rule to reject on one level, but 2 other levels are within 1 SD. Staff is just repeating the 1:3s. Is that the right approach? Also, do you apply rules across materials?" [Another question from the Cut Your QC webinar, now available as an on-demand recording: https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ  ]...

Continue reading
  343 Hits

Q: Should we change our 2 SD manufacturer range to a 3 SD manufacturer range?

"What is the solution if the quality control comes with 2 SD, do we need to change the product to a 3 SD control?" Another question from our Cut Your QC webinar, still available on demand at https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ I believe this question is referring to the range that comes with the control. The manufactu...

Continue reading
  265 Hits

Q: Will my staff get confused and frustrated by Westgard Sigma Rules?

Another question from the Cut Your QC webinar: "Most of my staff are generalists. It can be difficult for them to keep up with all of the details in every department. If we use different QC rules on each assay and instrument, how do we manage their frustration and confusion?" I get it. The prospect of every assay having a different set of QC rules ...

Continue reading
  324 Hits

Q: If controls are unstable, can they be trusted?

A follow-up question to one of the earlier Q&As: "If multiple laboratories observe a stability-related drift in a control material before the end of the manufacturer's claimed shelf life, shouldn't this trigger a reassessment of both the assigned target values and the validated stability claims by the manufacturer? "From my perspective, if real...

Continue reading
  282 Hits

Q: How do we account for issues with our controls?

Another question from the Cut Your QC webinar in September: "How do we account for issues with the controls themselves (i.e. real life stability is shorter than manufacturer's recommendation)?" Of course, this is the nightmare of the QC professional: if the controls introduce more variability, the signals from QC are not completely reflective of th...

Continue reading
  285 Hits

Q: Should I record my repeats?

Another question that came up in our Cut Your QC webinar last month: "If a lab is repeating QC until the QC "passes", must all the out-of-control QC be recorded in the LIS? Should they be marked as accepted or rejected?" If you take an action during troubleshooting, or during any step of your laboratory operations, you should record it. To selectiv...

Continue reading
  236 Hits

Q and A: Six Sigma Metrics and Accreditation

Another question from our Cut your QC webinar: Do accreditation bodies accept the Six Sigma metrics? Let's start out by being clear: Six Sigma is NOT required by CLIA, CAP, JCI, or any other regulatory or accreditation system in the world, as far as we know. It is an option, not a mandate. Thus, it's not something that has to be accepted. But it sh...

Continue reading
  376 Hits

Is Cutting Your QC driven by Risk? Or increasing it?

Another timely question from our Cut Your QC webinar 9/25/26: "What model can be used with planning QC based on the clinical risk to the patient, as considered ISO 15189? One of the challenges lies in selecting quality specifications – specifically, how to determine whether the approach is too lax or too strict." Risk-driven, Risk-focused, Risk-cen...

Continue reading
  242 Hits

Cut your QC Question: a 1-2s by the light of the moon

We got a lot of questions during our Cut Your QC webinar, too many to answer during the session. So here's one that we didn't get to "In the evening I run one level of QC and it fails 1:2s. Do I need to rerun?" This is a curious question. QC results are not different depending on the time of day. A 1:2s in the morning is the same as a 1:2s at night...

Continue reading
  272 Hits

What does "Cut Your QC" actually mean?

As we prepare for our Cutting Your QC webinar, we've gotten nearly two thousand registrations, along with a lot of questions. https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ The most common question we're getting is, What does "Cut your QC" mean? I believe this comes from a perspective of many labs, that QC cannot be reduced...

Continue reading
  963 Hits

AI is getting the Westgard Rules Rong...

Over the past few months, I've gained a bit of a reputation as a LinkedIn grouch. When I come across posts here that describe the Westgard Rules, analytical Sigma metrics, QC in general, I am quite direct about pointing out errors. In the flood of AI-generated images and flyers, there is currently an overwhelming trend to get things wrong. Over the...

Continue reading
  370 Hits

Cut Your QC - The caveats

More than 600 of your laboratory peers are already registered for Cut Your QC. https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ Now there are some qualifiers, of course. Not every lab nor every instrument is going to be able to reduce QC. A badly engineered instrument. Or a poorly maintained and deteriorating instrument. Or a poor...

Continue reading
  581 Hits

Diagnostics Peer-Group Sigma Study: Roche beats Siemens

An Indonesian study in Diagnostics attempted to make a comparison between Siemens and Roche instrumentation using Peer Group Sigma metrics. Out of 30 assays, Roche won 29 of 30. I can't say I wholeheartedly agree with the methodology of the study: peer group data can obscure what individual laboratories experience. The details can get lost in ...

Continue reading
  543 Hits

Six Sigma for Agriculture and Textile Chemical?

Like many authors, I keep track of where I'm cited. So I was a bit surprised to see I had been referenced in a paper in the Agricultural Association of Textile Chemical and Critical Reviews Journal: https://doi.org/10.21276/AATCCReview.2026.14.03.22 Not the usual journal I expect. But even more curious, the title of the paper is "An overview of qua...

Continue reading
  485 Hits

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input