A Six Sigma Series from Advance

Posted by Sten Westgard, MS

Last month, Advance for the Laboratory published a three part series on Six Sigma in the Clinical Laboratory, written by David Plaut, Nathalie Lepage, and Kim Przekop:

Part 1

Part 2

Part 3

While it's great that Advance has invested in in-depth coverage of the Six Sigma topic, unfortunately one of their examples in part 2 demonstrates a misunderstanding of the application.

See the mistake, after the jump...

-----

In the example, three different analytes are assessed, and a lot of confusion might result. Either there's a mistake in the table or there's a mistake in their calculations.

Figure 2

With the first example, hemoglobin, the mean of a method is 12.4 g/dL while the group mean is 12.3 g/dL, and the lab SD is 0.7 g/dL. That means the CV is (0.7/12.4) = 5.6% and the bias is (12.4-12.3/12.3) = 0.8%. Given that the CLIA requirement for TEa is 7%, this is not a Six Sigma method. Instead, the Sigma-metric is (7 - 0.8)/5.6 = only 1.1.

Now, if we believe the SD is being expressed in table as a percentage and not units (i.e. that "g/dL" in the second row left column is a mistake), the Sigma metric is now (7 - 0.8) / 0.7 = 8.8 Sigma. Unfortunately, this is a number not listed in the table. So we can't be sure this is what they did.

The example attempts to derive through a laborious method an estimate of the SD necessary to achieve 6 Sigma. The easiest way to do that is simply divide the TEa by 6:  7/6 = 1.16% or about 1.2%. In units at a level of 12.4 g/dL that works out to about 0.2 g/dL. The table lists 1.2 g/dL, which is six times the ideal SD.

Whenever you mix units and percentages, there's always a danger you make miscalculations. That's why we strongly encourage labs to convert everything into percentages, and then make the Sigma-metric calculation only in those percentages.

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

MLP Week: Bring Back the Bobble, Labs!
Free webinar: Simplify your IQCP
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Saturday, 14 November 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input