Posted by Sten Westgard, MS
Posted by Sten Westgard, MS
Posted by Sten Westgard, MS
"We have a department supervisor that instructs the techs to delete (not omit) qc values outside of three s.d..Is there a specific CLIA rule against this? Where I come from, deleting qc values is wrong."
What do you think? The answer, after the jump...
-----Posted by Sten Westgard, MS
Posted by Sten Westgard, MS
The third year of Medicare penalties has been meted out. This year hospitals were penalized for avoidable complications such as various types of infections, blood clots, bed sores, falls, and new to the list this year, the prevalence of two types of bacteria resistant to drugs.
Do you want to guess the percentage of injuries that impacted patients in US healthcare?
The answer, after the jump...
-----Posted by Sten Westgard, MS
Another question coming in from one of our website members:
Could you give me some suggestions on establishing QC ranges for unassayed chem. Controls?
We currently use [Brand X] unassayed chem controls (much less in cost than assayed controls) for some of our chemistry analytes.
For these controls, we are provided a “target mean” and a range.
Occasionally, our established mean (i.e. n=30) for a new lot is outside of the range provided for “target mean” provided by manufacturer.
Could you suggest guidelines for acceptance of lab established means for unassayed controls?
Obviously, I would like to know how far from the “target mean” could the new lab established mean be?
The answer, following the jump...
-----Posted by Sten Westgard, MS
We got the following question emailed to the website recently:
Greetings, I was [wondering] if the following case represents [violation] of 4 1s rule?
Control (A) point 1= EXCEEDS 2 SD /-2 SD lines >> warning
Control (A) point 2= WITHIN 2 SD/-2 SD lines >> accepted
Control (A) point 3= EXCEEDS 2 SD /-2 SD lines >> warning
Control (A) point 4= WITHIN 2 SD/-2 SD lines >> rejection
all the point[s] on the same side of the mean.
What do you think? The answer, after the jump
-----Posted by Sten Westgard, MS
Posted by Sten Westgard, MS