This table contains information on Clinical quality requirement that describe medically important changes in test values, or Decision Intervals (Dint) expressed as a percentage change at a certain Decision Level (Dint = change divided by decision level multiplied by 100 to give a percentage). The sources of the decision intervals presented here are the paper by Skendzel, Barnett, and Platt and a series of recommendations for lipid tests from the National Cholesterol Education Program (NCEP).
To make practical use of these clinical quality requirements, it is also necessary to have information about pre-analytical factors, particularly within-subject biological variation(swsub). The sources of estimates of within-subject biological variation presented here are summaries provided by Fraser and individual references from or cited by NCEP in their recommendations for lipid tests.
Clinical Quality Requirements | |||||
---|---|---|---|---|---|
Test | Decision Level | Dint(%) | Source | swsub(%) | Source |
AST | 30 IU.L 70 IU/L |
43.3% 28.6% |
Skendzel Skendzel |
10.8% | Fraser |
Bilirubin, total | 0.8 mg/dL 15.0 mg/dL |
75.0% 13.3% |
Skendzel Skendzel |
18.1% | Fraser |
Calcium | 8.5-9.0 mg/dL | 10/6% | Skendzel | 2.0% | Fraser |
Cholesterol, total | 210 mg/dL 200 mg/dL |
33.3% 20.0% |
Skendzel NCEP |
6.1% 6.5% |
Fraser Costongs |
Cholesterol, HDL | 60 mg/dL | 42.0% | NCEP | 7.5% | Warnick |
Cholesterol, LDL | 130 mg/dL | 23.1% | NCEP | 8.2% | Backorik |
Creatinine | 1.0 mg/dL 3.0 mg/dL |
50% 16.7% |
Skendzel Skendzel |
4.9% | Fraser |
Glucose | 100 mg/dL | 30.0% | Skendzel | 12.2% | Fraser |
Iron | 150 ug/dL | 33.3% | Skendzel | 19.8% | Fraser |
Potassium | 3.8 mmol/L | 10.5% | Skendzel | 4.9% | Fraser |
Sodium | 125 mmol/L 135mmol/L |
4.0% 4.4% |
Skendzel Skendzel |
0.8% | Fraser |
Thyroxine, total | 6.0 ug/L | 33.3% | Skendzel | 5.8% | Fraser |
Total protein | 7.0 g/dL | 21.4% | Skendzel | 3.3% | Fraser |
Triglyceride | 130 mg/dL | 46.2% 100% |
Skendzel NCEP |
19.8% 23.7% |
Fraser Stein |
Urea nitrogen | 18.0 mg/dL 40.0 mg/dL |
56.6% 20.0 |
Skendzel Skendzel |
12.2% | Fraser |
Hematocrit | 42 mg/dL | 11.9% | Skendzel | 2.5% | Fraser |
Hemoglobin | 15.0 g/dL | 8.0% | Skendzel | 2.4% | Fraser |
Leukocyte count | 5x109 cell/L 25x109 cell/L |
32% 28% |
Skendzel Skendzel |
15.6% | Fraser |
Prothrobin time | 12.0 sec 20.0 sec |
25.0% 25.0% |
Skendzel Skendzel |
2.3% | Dot |