Q: Should we change our 2 SD manufacturer range to a 3 SD manufacturer range?

"What is the solution if the quality control comes with 2 SD, do we need to change the product to a 3 SD control?"

Another question from our Cut Your QC webinar, still available on demand at

https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ

I believe this question is referring to the range that comes with the control. The manufacturer typically gives a range, meant to be a range of means (that is, you should find your mean inside this range), which some laboratories adopt as their actual control range.

When we talk about moving to 3 SD, this refers to the actual QC rules being used on the method. Instead of the typical 1:2s rule, a 6 sigma test can use a 1:3s rule. But this rule must be based on the actual mean and sd of the laboratory. It cannot be used on the manufacturer's range, because the SD given from the manufacturer does not represent the actual SD of the laboratory.

So a laboratory can use different (and better and fewer) QC rules when they apply Westgard Sigma Rules, but only if they use an honest foundation built with their own mean and own SD. You can't deploy Westgard Sigma Rules on someone else's mean or SD.

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Q: 1 level is out 3 sd, but the other 2 levels are...
Q: Will my staff get confused and frustrated by We...
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Thursday, 22 October 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input