Q: What if we can't Westgard Sigma Rules?

Another Question from the "Cut Your QC" webinar:

"If management is unwilling to adopt the Sigma-metric method, is there a better Westgard Rule than 1:2s to apply across the board to all analytes?"

If Westgard Sigma Rules aren't a possibility at your laboratory, there are still better ways to run QC using the 1:2s rule. Remember, with 2 controls, the 1:2s rule will generate 9% false rejection. Almost 1 of 10 runs is going to be out of control. So if your lab has you'll probably see alarms every day

If you can't - or won't - characterize the performance of the method, then applying a uniform set of modern Westgard Rules will still be an improvement over the 1:2s.

The 1:3s/2:2s/R:4s/4:1s/8:x with 2 controls, where the 1:3s/2:2s/R:4s can be interpreted in the first run (R=1), 2 runs are needed for the 4:1s interpretation, and 4 runs are needed for the 8:x interpretation. This multirule set will give about 3% false rejection, which is a 66.6% reduction from the 1:2s with 2 controls.

Even that simple step will provide significant savings, but only if you stop reacting to the 2sd outliers with repeats, recalls, etc. That may be the hardest part: getting staff that have been accustomed to repeating controls for years, even decades, to un-learn that bad habit.

Have a question? Send it to This email address is being protected from spambots. You need JavaScript enabled to view it.

Watch the recording of Cut Your QC here: https://us02web.zoom.us/webinar/register/WN_F5IRlqySRR6eRcrm9Zz9mQ passcode: 746694

[Available through the generous support of Thermo Fisher MAS controls and LabLink 360]

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Q & A: We don't run QC with a time machine
Q: For every method, instrument, reagent lot, a ne...
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Sunday, 08 November 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input